XP8500
and
Exp-A/D12
Technical Circuit Details
s
103
Settling times were also measured for other channels.
Reducing the size of the feedback capacitor reduced the settling time at the
expense of reduced noise filtering. When C10 (202X gain) was replaced
with a 150 pF capacitor, the time constant on this channel was reduced to
35 µs. The stability of the channels suffered disproportionately.
Summary of Contents for Exp-A/D12
Page 6: ...vi s Contents XP8500 and Exp A D12 Blank...
Page 10: ...XP8500 and Exp A D12 x s About This Manual Blank...
Page 11: ...XP8500...
Page 12: ...Blank...
Page 49: ...EXP A D12...
Page 50: ...Blank...
Page 58: ...Exp A D12 58 s Getting Started Blank...
Page 64: ...Exp A D12 64 s I O Configurations Blank...
Page 73: ...APPENDICES...
Page 74: ...Blank...
Page 86: ...XP8500 and Exp A D12 86 s PLCBus Blank...
Page 87: ...XP8500 and Exp A D12 Specifications s 87 APPENDIX B SPECIFICATIONS...
Page 92: ...XP8500 and Exp A D12 92 s Specifications Blank...
Page 97: ...XP8500 and Exp A D12 Simulated PLCBus Connection s 97 APPENDIX D SIMULATED PLCBUS CONNECTION...
Page 99: ...XP8500 and Exp A D12 Technical Circuit Details s 99 APPENDIX E TECHNICAL CIRCUIT DETAILS...
Page 104: ...XP8500 and Exp A D12 104 s Technical Circuit Details Blank...