2
2-3 DC millivolt (DCmV)
Range
Accuracy
Resolution
400mV
±(1.0%+10)
0.lmV
2-4 True effective value of AC millivolt (ACmV)
Range
Accuracy
Range scope
Resolution
400mV
±(3.0%+3)
40Hz-lkHz
0.lmV
2-5 True effective value of D\AC voltage (ACV)
Range
Accuracy
Frequency
scope
Resolution
4V
±(1.0%+3)
40Hz-1kHz
lmV
40V
10mV
400V
100mV
750V
1V
Input impedance: 400mV range > 40M
Ω
, and the other is 10 M
Ω
;
Overload protection: DC1000V or AC750V peak value;
Frequency response: The frequency scope for all ranges is 40Hz – 1kHz (applicable to standard sine wave and
triangle wave).
Display: True effective value (more than 200Hz for other waveforms, only for reference).
2-6 DC current (DCA)
Range
Accuracy
Resolution
400uA
±(1.5%+3)
0.luA
4000uA
luA
40mA
10uA
400mA
100uA
10A
10mA
Maximum voltage drop measured: Full range mA: 400mV; A: 100mV;
Maximum input current: 10A (not more than 10s);
Overload protection: 0.5A/250V fuse; 10A/250V fuse.
2-7 True effective value of AC current (ACA)
Range
Accuracy
Frequency
scope
Resolution
400uA
±(1.5%+3)
40Hz-lkHz
0. luA
4000uA
luA
40mA
10uA
400mA
100uA
10A
±(l,5%+3)
10mA
Maximum voltage drop measured: Full range mA: 400mV; A: 100mV;
Maximum input current: 10A (not more than 15s);
Overload protection: 0.5A/250V fuse; 10A/250V fuse.
Frequency response: The frequency scope for all grades of current is 40Hz-1kHz; (applicable to standard sine
wave and triangle wave; and that for other waveforms: more than 200Hz, only for reference).
2-8 Resistance (
Ω
)
Range
Accuracy
Resolution
400
Ω
±(0.5%+3)
0.1
Ω