Zutemer 17 User Manual Download Page 8

 

7

4-14 Automatic power off    

1. If  being left unused for 15min, the instrument will be powered off  automatically, and then enter into sleep 

state, but in 1min before power-off, the built-in buzzer will give a prompt by ringing for five times; later, the 

power supply may be restarted by pressing any key.   

2. Press “select” key before starting the machine may cancel the function of  automatic power-off.   
 

V. Maintenance of  the Instrument   

This instrument is precise, and users shall never change its circuit at random.   

1. Protect the instrument from water, dust and hurl-down.   

2. Never store or use the instrument in high-temperature, high-humidity, flammable, explosive and 

strong-magnetic-field environment;   

3. Please clean the outer surface of  the instrument with wet fabric and gentle detergent, instead of  grinding 

additives and strong solvents like alcohol, etc.; 

4. If  the instrument is left unused for long term, the battery shall be taken out in order to avoid that the 

instrument is eroded by battery weeping;   

5. Pay attention to the use of  battery, and replace the battery if  LCD displays the sign “

”;  

The steps are as follows:   

5-1 Screw out the screws for fastening the battery on back cover, and remove the battery door; 

5-2 Replace the 9V batteries with two new batteries. Any standard 9V battery is available, but it’s best to use 

alkalic batteries in order to extend the service life.   

5-3 Install the battery door and fasten the screws;   

6. Fuse replacement: The steps are the same as above. For replacement of  fuse, please use fuse of  the same 

specification and model.   

NOTES:  

1. Don’t connect the voltage higher than DC1000V or AC750V peak value;   

2. Don’t measure voltage value at AC, resistance, diode, and buzzer levels;   

3. Don’t use this instrument is the batteries are not installed or the back cover is not fastened;   

4. Please move the testing probes from tested points and shut down the machine before replacement of  

batteries or fuse.   
 

VI. Troubleshooting   

If  your instrument does not work normally, the following methods may help you solve common problems. If  

the troubles can not still be eliminated, please contact our maintenance center or dealer.   
 
Fault phenomena 

Position to be inspected and method 

No display 

The power supply is not connected; 
Holding switch; 
Replace the batteries.   

Appearance of  the sign 

 

Replace the batteries. 

No input of  current 

Replace the fuse. 

Big error in display   

Replace the batteries.   

 
Any amendment to the specifications won’t be notified;   

The contents mentioned herein are considered correct, and in case of  discovering any error and omission, 

etc., users shall please contact the manufacturer; 

The Company doesn’t assume responsibility for the accidents and hazards arising from users’ misoperation; 

The functions stated herein cannot be taken as reason for the product to be used for other purposes.   

 

 

 

 
 

www.zutemer.cz

    |   

[email protected]

   

 

 

 

Summary of Contents for 17

Page 1: ...Digital Multimeter ZUTEMER 17 User Manual www zutemer cz ...

Page 2: ... 4 Select correct function and range to avoid misoperation The instrument of this series has full range protection function but for the sake of safety you shall still pay attention to this point 5 Do not input a current higher than 10A when measuring the current 6 The safety sign indicates the existence of dangerous voltage indicates grounding indicates dual insulation indicates that operator must...

Page 3: ...nce 2 6 DC current DCA Range Accuracy Resolution 400uA 1 5 3 0 luA 4000uA luA 40mA 10uA 400mA 100uA 10A 10mA Maximum voltage drop measured Full range mA 400mV A 100mV Maximum input current 10A not more than 10s Overload protection 0 5A 250V fuse 10A 250V fuse 2 7 True effective value of AC current ACA Range Accuracy Frequency scope Resolution 400uA 1 5 3 40Hz lkHz 0 luA 4000uA luA 40mA 10uA 400mA ...

Page 4: ... protection DC or AC250V peak value Warning It s forbidden to input voltage value in this range for the sake of safety 2 11 Temperature Range Accuracy Resolution 20 1000 400 C 土 1 0 5 400 1 5 15 1 4 1832 752 1 0 5 752 F 1 5 15 1 Sensor K type nickel chromium nickel silicon Warning It s forbidden to input voltage value in this range for the sake of safety IV Application Method 4 1 Description of op...

Page 5: ...oint contacted by the red multimeter probe will be displayed on the screen NOTES 1 Under the mode of manual range if the LCD displays OL it indicates going beyond the range and RANGE key shall be rotated to a higher level 2 The voltage measured shall never exceed DC1000V or the instrument circuit will possibly be damaged 3 Never contact the high voltage circuit measured 4 3 DC millivolt voltage me...

Page 6: ...red multimeter probe into mA or 10A hole maximum 400mA or 10A 2 Rotate the functional switch to current level press SELECT key to select AC measurement mode and then connect the probe of the instrument to the tested circuit The current value tested will be displayed on the screen simultaneously NOTES 1 If having no idea about the scope of current measured users shall rotate the range switch to the...

Page 7: ... probe into VΩHz hole the polarity of red multimeter pole is 2 Rotate the functional switch to or level 3 Measurement in forward direction Connect the red black multimeter probe to the positive negative pole of the tested diode and the display unit will display the approximate value of the voltage drop in forward direction of the diode 4 Measurement in backward direction Connect the red black mult...

Page 8: ...lkalic batteries in order to extend the service life 5 3 Install the battery door and fasten the screws 6 Fuse replacement The steps are the same as above For replacement of fuse please use fuse of the same specification and model NOTES 1 Don t connect the voltage higher than DC1000V or AC750V peak value 2 Don t measure voltage value at AC resistance diode and buzzer levels 3 Don t use this instru...