background image

FV533  Rev. A   1/13/11
Page 2

PRIOR TO INSTALLATION

Prior to installing the ZTS6200 series flushometer valve,
install the items listed below:
• Fixture
• Drain line
• Water supply line

IMPORTANT:

• All Plumbing is to be installed in accordance with applicable

codes and regulations.

• Water supply lines must be sized to provide an adequate

volume of water for each fixture.

• Flush all water lines prior to operation (See Step 2).
• Dirt and debris can cause flush valve to run continuously.
• Sensor units should not be located across from each other or

in close proximity to highly reflective surfaces.

The ZTS is designed to operate with 40 to 116 psi static water pressure.  Protect the chrome or special finish of this flushometer.

Do not use toothed tools to install or service the valve.

1.) 

Install stop valve assembly 

(A)

 using proper size supply

escutcheon and sweat solder adapter kit if applicable.  Thread
sealing compounds should be used on male NPT threads
only. 

See Figure 1

5.) 

Prior to inserting the flush valve tailpiece 

(B)

 into stop valve

(A)

, be certain that the O-ring seal 

(C)

 is located in O-ring

seal groove at the end of the tailpiece and that the locking nut

(D) 

and locking snap ring 

(E)

 are located as shown.  Care

should be taken not to damage the O-ring when inserting
the tailpiece into the stop valve. If lubrication is needed, wet-
ting the O-ring with water will be sufficient. Tighten all con-
nections with a non-toothed wrench. 

See Figure 5

Figure 1

E

C

A

D

Before the supply water is turned on, be sure all stop valves  are
closed off tight. The stop valves can be opened and closed by
using the adjusting screw 

(S)

 located at the center of the stop

valve cap 

(T)

. Stop valve adjustments can only be made by us-

ing the adjusting screw 

(S)

It is not necessary to remove the

stop valve cap (T) when making adjustments. If for any rea-
son it becomes necessary to remove the stop valve cap (T),
be certain the water is shut off at the main supply valve. 

See

Figure 2

T

S

Figure 2

Figure 3

3.) 

Install battery into cap. *

Note: Do not remove lens label at

this time.

 This is to prevent the sensor from inadvertently

going into calibration mode.

4.) 

After battery is installed, connect solenoid to electronics and

place cap on valve body.

   

 *Note: Install cap with care and insure wires are not pinched

or damaged. This can cause the valve not to function prop-
erly.

Once cap is securely emplaced on the valve, install the set
screw to lock the cap on the valve. Tighten set screw with
allen wrench provided.

B

A

2.) 

When all stops are connected to the water supply and

    water pressure is available, it is recommended that the
    supply piping be flushed to remove dirt, metal chips, etc.,
     from system.

A. Before the valve is installed, open each stop fully for a brief

time and catch the water in a two gallon or larger bucket.  For
multiple installations, start with the stop valve closest to the
water supply and work toward the most remote valve.

B. Due to the small passages and orifices, it is not possible to

flush the piping through the low volume valve.

C. Once the lines are flushed, the valve can be installed.

Figure 4

Figure 5

Summary of Contents for AquaSense ZTS6200

Page 1: ...foranylabor chargesoranyloss injuryordamageswhatsoever includingincidentalorconsequentialdamages Thesoleandexclusiveremedy shallbelimitedtothereplacement ofthedefectivegoods Beforeinstallationanduse thepurchasershalldeterminethesuitabilityoftheproductforhisintendeduseandthepurchaserassumesallriskand liability whatever in connection therewith Where permitted by law the implied warranty of merchanta...

Page 2: ...e 1 E C A D Before the supply water is turned on be sure all stop valves are closed off tight The stop valves can be opened and closed by using the adjusting screw S located at the center of the stop valve cap T Stop valve adjustments can only be made by us ing the adjusting screw S It is not necessary to remove the stop valve cap T when making adjustments If for any rea son it becomes necessary t...

Page 3: ...ation will occur when the valve senses an object person The LED will blink once At that point the valve is charged When object leaves view of the sensor valve will activate An eight second sensing cycle will prevent the valve from flushing needlessly when someone walks by If special circumstances require adjustment of the sensing distance see Appendix A Changing Activation Distance See Figure 9 B ...

Page 4: ...battery to the sensor Note There is a 5 second window after connecting the battery in which the sensor range adjust mode can be accessed 4 Place your hand in front of the sensor at a distance of 2 to 4 within 5 seconds after battery is reconnected to enter sensor range adjust mode 5 Initially the red light in the sensor will flash slowly 3 times approximately 2 seconds per flash Keep your hand in ...

Page 5: ...or other objects 1 Connectors between the electronic unit and solenoid are disconnected 2 Debris or scale in solenoid 3 The water supply pressure is too high 1 Debris or scale in piston 2 Sensor is dirty or covered 3 Sensor is picking up reflections from mirror or another object Solution 1 Replace battery 1 Increase the range 2 Decrease the range 3 Replace battery 4 Eliminate cause of reflection 1...

Page 6: ...onnections and Spud Coupling Kits Product No Flush Connection and Spud Coupling Items 15 18 P6000 H P6003 H VacuumBreaker Repair Kit Items 11 12 P6000 B VacuumBreaker Tube P6000 A CP VacuumBreaker Tube Nut P6000 AA CP Control Stop Repair Kit and Parts Product No Control Stop Repair Kit for 1 and 3 4 P6000 C SD CP Includes Items 14 20 Seal Seat for 1 and 3 4 Includes Item23 P6000 D42 Sw eat Solder ...

Reviews: