![ZURN AquaSense ZEMS6142AV Installation, Operation, Maintenance And Parts Manual Download Page 6](http://html.mh-extra.com/html/zurn/aquasense-zems6142av/aquasense-zems6142av_installation-operation-maintenance-and-parts-manual_942606006.webp)
FV38-WCS • Page 6
Figure 12
Figure 13
STEP 10 – Test and Adjustment (Figure 13)
(Also see “Sensor Features” on Page 7.)
Zurn Automatic Sensors are preset at the factory and provide
the user with selectable options as follows:
Switch 1 – Automatic Flush (Factory setting is on.)
Switch 2 – Red/Green LED (Factory setting is on.)
Switch 3 – Courtesy Flush (Factory setting is on.)
• Red/Green LED –
With power applied the red light will
illuminate when an object is detected. The green light will
illuminate after the object has been detected for a minimum
of 8 seconds then leaves. The green light indicates the
flushing sequence.
• Range Adjustments –
The Zurn Automatic Sensors are
factory set to operate at a range of 50". This range should
be satisfactory for most installations. If the range is too short
(i.e., not detecting the user) or too long, adjust the range.
•
To Make A Range Adjustment –
Remove the rubber push
caps. The range adjust ment is shown in Figure 13. Using a
small screwdriver, turn the range adjustment clockwise to
increase range (maximum range is 60") or counterclockwise
to decrease range (minmum range is 12").
CAUTION:
Range adjustment rotates only 1/2 turn total from
min. to max. Do not exceed this or sensor will be damaged.
Stand in front of the Zurn Automatic Sensor at the desired distance;
the red light will illuminate when you are within range.
Repeat adjustment procedure until the desired range is obtained.
• Switch Settings
(Figure 13)
The three switches control the following options. They can be
activated or deactivated by means of the switches (down is
on, up is off).
Switch 1 – Automatic Flush –
The unit will auto matically
flush 24 hours after last user.
Switch 2 – Red/Green LED –
When an object is detected
the Red LED illuminates. The Green LED illuminates while
flushing.
Switch 3 – Courtesy Flush –
When an object has been
detected for two seconds, the unit flushes and goes into the
normal mode. This is repeatable after 90 seconds.
NOTE:
After the options have been selected the reset button
must
be pushed for sensor to accept selections. Make sure
rubber push caps are securely installed after making adjust-
ments to protect sensor from moisture.
STEP 11 – Adjust Control Stop
(Figures 6 and 12)
Adjust the control stop to meet the flow rate required for proper
cleaning of the fixture. Open control stop counter clockwise one-
half turn from closed position. Activate Flushometer simulating a
user. Adjust the control stop after each flush until the rate of
flow delivered properly cleanses the fixture.
RUBBER
PUSH CAPS
RESET BUTTON
RANGE
ADJUSTMENT
SELECTION
SWITCHES
(SHOWN IN
FACTORY
POSITION)
SENSOR
COVER
GREEN
BLUE (–)
BROWN (+)
VALVE BODY COVER
INTERNAL PLASTIC
COVER
TRIP MECHANISM
CONTROL STOP
INTERNAL KIT