Installation Description
17
/
44
Do not install the vein illuminator at a place where the power
plug is difficult to be disconnected from the power socket.
5.4
Checking the Device Accuracy
Since the vein illuminator is a precise device, before powering
on it for the first time or after transporting it, the operator must check
the accuracy of the device using an accuracy test card.
Place the accuracy test card in the projection area of the device,
and check the positional deviation of the projection image and test
card image at the optimal focus position (the projection hole is
210±30mm away from the test card). If the deviation is smaller than
1mm, the device can be used properly. If the deviation is larger than
1mm, stop using the device and contact the local distributor or the
after-sales service department of MEDCAPTAIN immediately.