Advanced programming
ZOLL Lifecor internal use only
Page B-15
Default settings
All settings return to their default values when you set up a new patient.
Menu option
Default setting
VT rate threshold
150 beats per minute
VF rate threshold
200 beats per minute
Secondary language
Off
VT response time
60 seconds
VF response time
25 seconds
MD notification option
Off
Patient sleep interval
00:00 (midnight) and 06:00 (6:00 a.m.)
Response time extension
0 seconds
Pulse energy
150 joules for all 5 shocks
Post treatment message
Blank
Patient dial prefix
Determined by device provider
Patient dialing number
Determined by device provider
Patient dialer mode
Tone
Hospital dial prefix
Blank or determined by ZOLL Lifecor
Hospital dialing number
Blank or determined by ZOLL Lifecor
Hospital dialer mode
Tone
Clinical center code
Determined by ZOLL Lifecor
Clinical center name
Determined by ZOLL Lifecor
Time zone
Select from options determined by country
Daylight savings time (DST)
Default determined by country
Connection method
Connection method most likely used in country
Phone number build code
Defined by country
Modem type
Defined by country
Modem initialization string override
Blank, use only as directed by ZOLL Lifecor
Primary dialing number override
Blank, use only as directed by ZOLL Lifecor
Primary login override
Blank, use only as directed by ZOLL Lifecor
Secondary dialing override
Blank, use only as directed by ZOLL Lifecor
Secondary login override
Blank, use only as directed by ZOLL Lifecor