Troubleshooting Guidelines
9650-1210-01 Rev. T
13-7
AC Charger
Symptom
Recommended Action
The green and orange-yellow CHARGER ON indicators
illuminate alternately.
•
Verify battery is installed.
•
Turn unit ON to identify the fault condition.
•
Replace battery pack with a fully charged battery pack.
•
If problem persists, replace battery pack, unplug device
from AC mains and plug device back into AC mains.
Unit displays the
LOW BATTERY
message when it is
plugged into AC mains.
•
Replace battery pack with a fully charged battery pack.
•
Unplug device from AC mains and plug device back into
AC mains.
•
Verify
AC mains is working properly.
None of the CHARGER ON indicators are illuminated
when the device is plugged into AC mains.
•
Unplug device from AC mains and plug device back into
AC mains.
•
Verify
AC mains is working properly.
Summary of Contents for E Series
Page 2: ...9650 1210 01 Rev T E Series Operator s Guide...
Page 9: ...This page intentionally left blank...
Page 37: ...E Series Operator s Guide This page intentionally left blank...
Page 47: ...E Series Operator s Guide This page intentionally left blank...
Page 53: ...E Series Operator s Guide This page intentionally left blank...
Page 57: ...E Series Operator s Guide This page intentionally left blank...
Page 61: ...E Series Operator s Guide This page intentionally left blank...
Page 67: ...E Series Operator s Guide This page intentionally left blank...
Page 79: ...E Series Operator s Guide This page intentionally left blank...
Page 99: ...E Series Operator s Guide This page intentionally left blank...
Page 127: ...E Series Operator s Guide This page intentionally left blank...