Operator’s Shift Checklist for E Series Products (Semiautomatic)
1. Condition
2. Multi-function Pads
3. Paddles (if applicable)
4. Inspect cables for cracks, frays, broken wires, connector pins
5. Batteries
6. Disposable supplies
7. Memory Card
8. Operational checks
9. Please check the appropriate box after each use of this checklist.
Recommended checks and procedures to be performed at the start of each shift. For more detailed information, see
the E Series Operator’s Guide.
Date
_____________________________
Location
__________________________
Unit Serial Number
______________________
1st
Shift
2nd
Shift
3rd
Shift
Remarks
Unit clean, no spills, clear of objects on top, case intact
1 set preconnected / 1 spare (check expiration dates)
Paddles clean, free of gel and contaminants, not pitted
Release from paddle wells easily
A. ECG electrode cable, connector
B. Defibrillator paddle cables, paddle handle
C. Multi-function cable, connector
A. Fully charged battery in unit
B. Fully charged spare battery available
A. Electrode gel or gel patches
B. MFE Pads in
sealed
pouches — 2 sets
C. ECG electrodes
D. Recorder paper
E. Alcohol wipes
F. Razors
Empty memory card install in unit (if applicable)
A. Power On Sequence
Turn unit to ON, 4-beep tone heard
B. Defibrillator
Multi-function cable connected to test connector:
CHECK PADS
displays.
Press
ANALYZE
button, unit charges to 30 J
Press and hold
SHOCK
button,
TEST OK
is displayed and printed
Attach MFC to ECG Simulator, set to VF
Verify
CHECK PATIENT
prompt is displayed
Press ANALYZE. Verify unit charges to 120 J
Press SHOCK, verify shock was delivered.
C. Paddles (if applicable) (Manual mode only)
Paddles in holders: Set defib energy level to 30 joules, and simultaneously
press and hold both defib
discharge
buttons;
TEST OK
message on recorder.
D. Pacer Operation (Manual mode only)
Multi-function cable not connected to Test Connector
Turn to PACER, set pacer rate to 150 ppm, press RECORDER button
Pacer pulses occur ever 2 large divisions (10 small divisions)
Press
4:1
button, pulses occur every 8 large divisions
Set PACER OUTPUT to 0 mA, no
CHECK PADS
prompt
Set PACER OUTPUT to 16 mA,
CHECK PADS
prompt and alarm
Reconnect Multi-function cable to test connector.
Press Clear Pace Alarm softkey;
CHECK PADS
prompt disappears and Pace
alarm stops.
E. Recorder
Press RECORDER button; recorder runs. Press again; recorder stops.
Inspect recorder printout
Signatures
No action required
1st ___________________
2nd __________________
3rd ___________________
Minor problem(s) corrected
Disposable supplies replaced
Major problem(s) identified — UNIT OUT OF SERVICE
Summary of Contents for E Series
Page 2: ...9650 1210 01 Rev T E Series Operator s Guide...
Page 9: ...This page intentionally left blank...
Page 37: ...E Series Operator s Guide This page intentionally left blank...
Page 47: ...E Series Operator s Guide This page intentionally left blank...
Page 53: ...E Series Operator s Guide This page intentionally left blank...
Page 57: ...E Series Operator s Guide This page intentionally left blank...
Page 61: ...E Series Operator s Guide This page intentionally left blank...
Page 67: ...E Series Operator s Guide This page intentionally left blank...
Page 79: ...E Series Operator s Guide This page intentionally left blank...
Page 99: ...E Series Operator s Guide This page intentionally left blank...
Page 127: ...E Series Operator s Guide This page intentionally left blank...