General Maintenance
9650-1210-01 Rev. T
11-3
5. When charge ready tone sounds, use the defibrillator
ENERGY SELECT
buttons on either the sternum
paddle or defibrillator front panel to change the
selected energy to 20 joules.
6. Verify that the unit disarms itself.
7. Use the defibrillator
ENERGY SELECT
buttons on
either the sternum paddle or defibrillator front panel to
change the selected energy back to 30 joules.
MFC Setup:
1. Plug the MFC cable into the unit, making sure it is not
plugged into the test connector.
2. Set the selector switch on the front panel to DEFIB
and select 30 J.
3. Verify that the unit alternately displays the
CHECK
PADS
and
POOR PAD CONTACT
messages.
4. Insert the end of the Multi-Function Cable into its test
connector (Attached to the Multi-function cable), as
shown below.
5. Verify that the unit displays the
DEFIB PAD SHORT
message.
Energy Delivery Test
1. Press the
CHARGE
button on the front panel.
2. Wait for the charge ready tone to sound and verify
that the energy ready value displayed on the monitor
registers 30 joules (
DEFIB 30J READY
).
3. Press and hold the
SHOCK
button on the front panel
of the defibrillator (or the shock buttons on the
paddles) until discharge occurs.
4. Verify that the units displays the
TEST OK
message
and prints a stripchart of the event, noting the energy
delivered and impedance data.
Note:
If
TEST FAILED
appears, contact the ZOLL
Technical Service Department immediately.
Testing the Pacer (Pacer Version Only)
Perform these tests on all pacer-equipped units
periodically.
1. Connect the ECG leads and Multi-function cable to
the simulator.
2. Turn the selector switch to PACER.
3. Turn the pacer rate control to 150 ppm.
4. Press the
RECORDER
button to generate a strip,
and then again to stop printing.
5. Verify that the pacing stimulus markers (
) occur
approximately every 10 small divisions (2 large
divisions or 1 cm).
6. Press the
RECORDER
button to generate another
stripchart, then press and hold the
4:1
button on the
front panel.
7. Press the
RECORDER
button again to stop printing.
8. Verify that the frequency of the markers decreases (8
large divisions or 4 cm between each marker).
9. Turn the PACER OUTPUT
control to 0 mA.
10. Verify there are no
CHECK PADS
or
POOR PAD
CONTACT
messages.
11. Disconnect MFE Pads or paddles from the
Multi-Function cable, and slowly turn the knob up to
change the pacer output to 16 mA.
12. Verify that the Pace alarm sounds and a
CLEAR
PACE ALARM
prompt flashes.
13. Connect the Multi-Function cable to the test
connector, and press the
Clear Pace Alarm
softkey.
14. Verify that the unit Pace alarm stops and that the unit
removes error messages from the LCD.
0!#%2
2!4%
PPM
0!#%2
/54054
M!
Summary of Contents for E Series
Page 2: ...9650 1210 01 Rev T E Series Operator s Guide...
Page 9: ...This page intentionally left blank...
Page 37: ...E Series Operator s Guide This page intentionally left blank...
Page 47: ...E Series Operator s Guide This page intentionally left blank...
Page 53: ...E Series Operator s Guide This page intentionally left blank...
Page 57: ...E Series Operator s Guide This page intentionally left blank...
Page 61: ...E Series Operator s Guide This page intentionally left blank...
Page 67: ...E Series Operator s Guide This page intentionally left blank...
Page 79: ...E Series Operator s Guide This page intentionally left blank...
Page 99: ...E Series Operator s Guide This page intentionally left blank...
Page 127: ...E Series Operator s Guide This page intentionally left blank...