SpO2 Parameter Values
Pressing the
SpO2
parameter button highlights the Low SpO
2
Alarm Limit and enables its
value to be changed. The default low SpO
2
value at start up is 94%. The SpO2 parameter uses
the same Context Menu as the HR parameter.
The following table gives the options and ranges for the SpO2 parameter:
Parameter Window
Options /Range
Availability/Notes
Models
Primary Value
SpO2
%
84 to 100
Measurement
All
Secondary
Values
Not Applicable
Alarm Limits
Low Limit
Measured
Values
Pleth
Waveform
86 to 99, Off
SpO2
Context Menu
(note same
as HR Context Menu)
Pulse Ox
Default :
Standby
Standby, Off,
On
All
Fast SAT
Default : Off
Off/On
Fast SAT enables rapid tracking of
arterial oxygen saturation changes by
minimizing the averaging. This mode is
clinically applicable during procedures
when detecting rapid changes in SpO
2
is paramount such as induction,
intubation, and sleep studies.
Sensitivity
Norm
Max
Norm
adjusts the pleth signal
sensitivity.
Max
interprets and displays
data for even the weakest of signals.
Max
is recommended during
procedures and when clinician and
patient contact is continuous.
APOD
Off
Off, On
When on, this mode improves detection
of the "probe off patient" condition, but
reduces the ability to acquire a reading
on patients of low perfusion.
Averaging
8 Seconds
2 to 4, 4 to 6,
8, 10, 12, 16
Seconds
Adjusts the SpO2 and HR averaging
durations.