Installing a Battery Pack
The AED Pro unit accepts the following types of battery packs:
•
Rechargeable PD 4410-series battery packs
•
Disposable sealed lithium manganese dioxide battery packs
•
SurePower rechargeable lithium ion battery packs
When the unit displays the message
CHANGE BATTERY
, immediately replace the battery with
a fully charged battery pack.
Before you begin
Be sure the unit is turned off.
Be sure the battery pack to be installed is fully charged.
WARNING!
Do not use a rechargeable battery pack if the unit’s standby period will exceed 90 days.
Procedure
To install or replace a battery pack:
Step
Action
Notes
1
Pull the bottom edge of the battery
compartment latch outward.
When released, the latch slides upward.
2
Pull the top edge of the latch toward you
and then downward.
3
Lift the edge of the battery compartment
cover.
ZM040004A
ZM040004A