Applying Defibrillation Electrode Pads
9650-0350-01 Rev. J
ZOLL AED Pro Operator’s Guide
3–7
Applying Defibrillation Electrode Pads — Adult stat•padz II
WARNING!
ZOLL stat•padz II electrodes are for adult patients only; do not use them on patients
under 8 years of age.
To apply
stat•padz
II
defibrillation electrode pads:
Step
Action
1
Tear open the electrode package and unfold the inner
package to expose the electrodes.
2
Remove the square electrode from its backing material,
and place it on the patient’s upper right chest as shown.
3
Place your hand on the electrode edge. Using your
other hand, gently roll the electrode onto the patient’s
chest, pushing out air from beneath the electrode as
you go.
4
Remove the round electrode from its backing material,
and place it on the patient’s lower left chest as shown.
For female
patients, place the
electrode under
the patient’s left
breast.
5
Place your hand on the electrode’s edge. Using your
other hand, roll the electrode onto the patient’s skin,
pushing out air from beneath the electrode as you go.
6
Follow the AED Pro prompts.