C
HAPTER
2 T
ROUBLESHOOTING
2–4
www.zoll.com
9650-0308-01 Rev. A
AED Plus Error Log Messages
The AED Plus unit records internal errors in an error log, which you can retrieve by using a
Windows-based personal computer with ZOLL Administration Software (ZAS).
To retrieve the error log, follow these steps:
Note:
After saving the error log file on the computer, you can use ZAS to clear the error log
on the AED Plus unit.
Table 2-2 lists the critical error log messages.
Important:
If any of these messages appears in the error log, contact ZOLL Technical Service
as described on page vii.
Step
Action
1
Turn off the AED Plus unit.
2
Start the ZOLL Administration Software (ZAS) on the computer.
3
Position the AED Plus unit so that its IrDA port has a clear line-of-sight with the computer’s
IrDA port and the ports are about 10 to 18 inches (25 to 45 cm) apart.
4
Press and hold the unit’s
ON/OFF
button for at least 5 seconds to start the unit in
nonrescue mode.
5
After the unit issues the message
COMMUNICATIONS ESTABLISHED
, refer to the ZAS
help for instructions on saving the unit’s history file and viewing the included error log.
Table 2-2. AED Plus Error Log Messages
Error
No.
Error Message
1
The patient relay is not working properly
2
Attempting to charge the capacitor while the SAFELINE in the CPLD is set - should not
charge
3
The capacitor takes too long to charge
4
The H-bridge test failed while the bridge was disabled. There should have been 0 current
measured.
5
VCAP1 and VCAP2 are not within a defined range of each other
6
The H-bridge test failed while the bridge was enabled - an invalid current reading was
taken during the positive test on the H-bridge
7
The H-bridge test failed while the bridge was enabled - an invalid current reading was
taken during the negative test on the H-bridge
8
Defib did not charge during the power-on charge test
9
The Defib task is not receiving A2D vCap1 event samples - needed for charging
10
FnSafety has determined that the charge has dropped below a specified value during the
holding of charge.
11
The capacitor failed to charge
12
The capacitor over-charged
13
The capacitor contained an unsafe voltage in an idle state
Summary of Contents for Aed Plus
Page 1: ...Service Manual 9650 0308 01 Rev A March 2005...
Page 4: ...TABLE OF CONTENTS ii www zoll com 9650 0308 01 Rev A...
Page 10: ...PREFACE viii www zoll com 9650 0308 01 Rev A...
Page 20: ...CHAPTER 1 MAINTENANCE TESTS 1 10 www zoll com 9650 0308 01 Rev A...
Page 26: ...CHAPTER 2 TROUBLESHOOTING 2 6 www zoll com 9650 0308 01 Rev A...