Manual Self-Test
9650-0308-01 Rev. A
ZOLL AED Plus Service Manual
1–9
3.0 Manual Self-Test
Use this test to initiate the unit’s self-diagnostic procedure, which verifies the following:
•
Battery usage since last reset.
•
Defibrillation electrodes are properly connected.
•
ECG signal acquisition and processing circuitry is functional.
•
Defibrillator charge and discharge circuitry (operational at 2 joules) is functional.
•
Microprocessor hardware and software are functional.
•
CPR monitoring circuitry and compression depth sensor are functional (when
CPR-D•padz
are attached).
•
Audio output circuitry is functional.
•
Display screen (if present) is functional.
Tools Needed
•
None.
Test Setup
•
Turn off the AED Plus unit.
•
If they are not already in place, preconnect defibrillation electrodes to the unit.
Test Procedure
Test
Action
Expected Result(s)
Pass
Fail
3.1
Press and hold the
ON/OFF
button for at least 5 seconds.
•
Within 10 seconds after the unit is
turned on, the Status indicator
displays a green check.
•
On the top panel, the indicator
lights for the rescue actions
illuminate in sequence.
!
UNIT OK
!
!
If an IrDA connection is established, turn off the unit;
otherwise, the unit shuts down automatically.