Xenon lamp
Light Source Housing
6
Zolix Instruments Co., Ltd.
http://www.zolix.com.cn E-mail:[email protected]
make the cathode of lamp locate the center of reflector and make the screw tight.
3.3 Light source housing operation guide and safety precautions
3.3.1 Adjustment before the xenon lamp is lighted
1) Adjust 3 reflector turret at the same time, make the
diatance between light cup underside and adjusting
bracket base is about 3mm
2) Reove collimating lens barrel, then
adjust shift knob
and high knob of lamp, to make the electrode and the
electrode imaging as shown in the picture on the
right:
the diatance between cathode and cathode imaging is
about 2mm.
3)
After adjustment, restore the collimating lens barrel
and side plate in place.
Electrode
Electrode
imaging
2mm