Voltage indicator: output voltage display.
Current indicator: output current display.
Starting switch: the machine is in standby state after power on. After the switch is pressed, the
switch glows red, the instrument starts and glows, and the light fan starts; once the switch is
pressed again, the output is turned off, and the cooling fan is turned off after 5 minutes. When 5
minutes later, the fan is still working, the switch flashes, press, will not start the light source
again.
(
See below power internal and external control switch settings)
Set / actual current switch: press the button to set the output current of the observed
current meter; for the buttons up to show the actual current output in the current
indicator.
Current in / out selection: control output for local settings; press for external 0~5V
control output.
Voltage input: an external 0~5V voltage signal to control the output current of the
size of the input interface.
Current regulation knob: by adjusting the knob and setting button, the current value
is displayed on the current display table.
Front output interface: if the light source is connected to the front panel interface, the
1 and 2 pin interfaces of the connector of the rear panel must be short circuited,
otherwise the power supply will be protected.
Adjusting knob
:
Fan
:
The working voltage (0-24V) of the light source fan can be adjusted by this knob
Max I
:
The maximum output current can be adjusted by this adjustment knob
Min I
:
The minimum output current can be adjusted by this adjustment knob
V-adj
:
The voltmeter can adjust the display calibration of the voltmeter
I-adj
:
The display calibration of ammeter can be adjusted by this knob
Indicator light
:
Error-P
:
When the light is bright, the power supply shell is opened illegally, the starting switch is
invalid, the light source is closed, the light source fan works continuously, and the power supply
is released after the fault is eliminated
Error-L
:
When the light is bright, the light source chamber is illegally opened, the starting switch
is invalid and the light source is turned off; after the fault is removed, the power is re energized to
relieve the alarm
Power
:
The lamp is connected in parallel with the starting switch light. When the lamp continues