24
24
CHAPTER 4
INSTALLATION / FIRST START-UP
SETTING UP AND INSTALLATION
Important instructions for setting up the system
The Zirkonofen 600/V3 sintering furnace has to be operated under the following conditions only:
- Up to 2000 m above sea level.
- Ambient temperature range from 15 °C to 30 °C.
- Do not expose the system to direct solar radiation.
- Maximum relative air humidity of 75 % at 30 °C.
- Do not place it near an air conditioner. Avoid cold draughts.
- The Zirkonofen 600/V3 sintering furnace has to be moved and put down on surfaces with
suitable stability, in a dry place and in a non-corrosive environment.
- Stable, non-slip installation surface, which can bear a minimum weight of 200 kg.
- Safety distance of min. 2 m to combustible material.
- Safety distance of min. 2 m to water.
- Minimum distance of min. 1.5 m to all sides (for cooling air, access).
-
2NCEGVJGVCDNGQPCƀCVITQWPFCPFCNKIPVJGYQTMUWTHCEGJQTK\QPVCNN[
- Provide for adequate work height (consider the height of the front glass in raised position).
- Work and danger area must be easily visible and lit (at least 500 lux).
- Make sure that the set position of the sintering furnace during operation allows easy removal of
the power plug in case of emergency.
-
#PGNGEVTKECNUQEMGVKUTGSWKTGFYKVJCRQYGTQWVRWVQH8CUſZGFEQPPGEVKQP.C[
connection cable protected with strain relief. The socket and mains fuse must be up to the valid
national standards. The socket and the earthing wire must be inspected and tested before starting
the furnace.
-
6JGRQYGTUWRRN[OWUVDGEQPUVCPV8QNVCIGƀWEVWCVKQPUOKIJVFCOCIGVJGGSWKROGPVŏUEQORQPGPVU
- Power supply must be installed and attached by an expert.
The equipment must be easily accessible from all sides for operation and possible component
exchange.
- The system must not be connected to the electric circuit during the installation phase:
danger of electric shock.
-
0QQDLGEVUCTGVQDGRNCEGFQPVQRQHQTDGUKFGVJG<KTMQPQHGP8UKPVGTKPIHWTPCEGſTGJC\CTF
Connection to the power supply system via residual-current circuit breaker:
-
6JGFGXKEGUCTGGSWKRRGFYKVJCPKPVGITCVGF'/%OCKPUſNVGTVQOGGVVJGEWTTGPV'/%
standard 2004/108 EC.
Summary of Contents for zirkonofen 600/V3
Page 1: ...zirkonofen 6oo V3 operator s and maintenance manual...
Page 3: ...03 03...
Page 7: ...07 07...
Page 10: ...10 10...
Page 18: ...18 18 CHAPTER SAFETY RESIDUAL RISKS...
Page 19: ...19 19 CHAPTER 2 COMPONENTS COMPONENTS...
Page 22: ...22 22 CHAPTER 3 TECHNICAL DATA...
Page 28: ...28 28 CHAPTER 5 OPERATION...
Page 30: ...30 30 CHAPTER 6 SINTERING PROGRAMMES...
Page 39: ...39 39...