17/40
2022.12.13 (Rev. 0)
[REMIFU-ZTN(EN)(NM)]
ZTone Instructions For Use
3.4.4 Connection of Transducer
❏
The transducer is detachable from the main body. Before using the product, turn the lever
to connect the transducer and the body.
Figure 4. Transducer connection
3.4.5 Transducer
❏
The ZTone is equipment that generates a magnetic field by applying a strong current
to a transducer. As heat is generated from the transducer due to the strong current
used to create the magnetic field, cooling is performed by circulating the cooling oil
inside the transducer. If the transducer is used while in a tilted position, the cooling
oil will not reach all sections of the transducer, and injury or stoppage of operation
due to overheating of the transducer can result. For this reason, it is recommended
to place the transducer in a horizontal position as much as possible.
Caution
Maintaining the transducer in a horizontal position is
recommended.