Manufacturer’s declaration of
Electromagnetic Compatibility
14
Page 30
Table 1
Guidance and Manufacturing Declaration- Electromagnetic Emissions
The device
Zwave
Pro
is intended for use in the electromagnetic environment specified below. The customer or user
of the device
Zwave
Pro
should ensure that it is used in such environment.
Emission Measurement
Compliance
Electromagnetic Environment-Guidelines
RF Emissions in accordance with
CISPR 11
Group 1
The device
ZWave
Pro
must emit electromagnetic
energy in order to ensure its intended function.
Nearby electronic equipment may be affected.
RF Emissions in accordance with
CISPR 11
Class
A
The device
ZWave
Pro
is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public supply
network that also supplies buildings used for
domestic purpose.
Emissions of Harmonics in
accordance with IEC 61000-3-2
Class
A
Emissions of voltage fluctuations/
flickers in accordance with IEC
61000-3-3
Compliant
Table 2
Guidance and Manufacturing Declaration- Electromagnetic Immunity
The device
Zwave
Pro
is intended for use in the electromagnetic environment specified below. The customer or user
of the device
Zwave
Pro
should ensure that it is used in such environment.
Immunity
Tests
IEC 60601- Test Level
Compliance Level
Electromagnetic Environment -
Guidelines
Electrostatic
Discharge
(ESD) in
accordance
with IEC
61000-4-2
± 8 kV Contact Discharge
± 2 kV, ± 4 kV, ± 8 kV,
± 15 kV Air Discharge
± 8 kV Contact Discharge
± 2 kV, ± 4 kV, ± 8 kV,
± 15 kV Air Discharge
Floors should be made from wood,
concrete or ceramic tiles. If floors are
covered with synthetic material, the
relative humidity must be at least 30 %
Electrical
fast
transient/
burst in
accordance
with IEC
61000-4-4
± 2 kV
100 kHz repetition
frequency
± 2 kV
100 kHz repetition frequency
The supply voltage quality must
correspond to that of a typical
commercial or hospital
environment.
Surges in
accordance
with IEC
61000-4-5
-Line-to-
Line-
± 0,5 kV, ± 1 kV
± 0,5 kV, ± 1 kV
Surges in
accordance
with IEC
61000-4-5
-Line-to-
Earth-
± 0,5 kV, ± 1 kV, ± 2 kV
± 0,5 kV, ± 1 kV, ± 2 kV