Operation Instructions
7.4 Display and Buttons
7
Page 13
Description of the
display elements and
buttons
(1) Energy
Displays the set pulse energy. When the application is active, the bar graph is filled in.
The pulse energy can be set before and during the delivery of pulses. The pulse
energy is adjustable between 60–185 mJ in 10 mJ steps.
(2) Frequency
(3) Choice of
Handpiece
Displays the set frequency.
Frequency range: 1–22 Hz, adjustable via the right controller. The maximum
selectable frequency depends on the energy level selected.
Connection of one handpiece:
If only one handpiece is connected, the window selection displays the channel of the
connected handpiece. The handpiece is activated automatically regardless of the
connected channel.
Connection of two handpieces:
When connecting two handpieces, the handpiece at channel I is activated first.
The desired handpiece is activated by selecting directly the handpiece in the selection
window.
The activated handpiece is presented by the orange applicator head.
(4) Mode
Displays the selected operating mode. When activating the “mode” window, the
selection menu appears with the operating modes:
Series pulses, burst 4 pulses, burst 8 pulses, burst 12 pulses.
The desired mode of operation is selected directly in the corresponding row.
1
3
2
4
5
6
7
9
8
10