Instructions for use

Z

Field 

Dual  (MFG-03)

EN

Summary of Contents for Z Field Dual

Page 1: ...Instructions for use ZFieldDual MFG 03 EN...

Page 2: ......

Page 3: ...Illustrations Front of the device Fig 1 Front of the device 1 Applicator left side 2 Applicator right side 3 Display 4 Central control knob 5 Control unit 6 Swiveling castors 1 2 3 4 5 6...

Page 4: ...Illustrations Rear of the device Fig 2 Rear of the device 7 Connections for applicator hose 8 Ventilation grid 9 Mains connection 10 On off switch 11 Identification plate 10 11 7 8 9...

Page 5: ...Remaining time 13 Total operating time 14 Configuration icon 15 Intensity in 16 Start stop 17 Channel selection channel 1 and or channel 2 18 Applied program 19 Mode selection Programs Region Manual M...

Page 6: ...Illustrations Accessories 21 Mains cable 22 Fixingbelt...

Page 7: ...arns of caution regarding possible damage to the device Applied part of type BF Consult instructions for use Follow the instructions for use No access for people with pacemakers Do not push the device...

Page 8: ...batteries and accumulators This product must not be disposed of with household waste Washing maximum 40 C Do not bleach Do not dry clean Do not tumble dry Do not iron Do not stack Keep dry Fragile ha...

Page 9: ...ion of symbols Page 1 Indications Contraindications Side Effects 1 2 Warnings 3 3 Z Field Dual overview Intended purpose 5 4 Setting up the device 6 5 Settings 7 6 Operating instructions 9 7 Technical...

Page 10: ...s Declaration on Electromagnetic Compatibility EMC 26 16 Material Safety Data Sheet 30 Valid for Z Field Dual MFG 03 devices These operating instructions form part of the device They must be stored t...

Page 11: ...fied with Z Field Dual treatments therefore we recommend not undergoing treatment during this time of the month Other contraindications to treatment include the following Fever Application over menstr...

Page 12: ...nt experiences abnormal symptoms the operator must stop treatment immediately and contact the appropriate physician The adverse effects may include but are not limited to Muscular pain Temporary muscl...

Page 13: ...castors are not in transport position Sitting on the device is prohibited Before use make sure that the device is supplied with power via a properly earthed plug with a safety plug electrical install...

Page 14: ...ct your service engineer In case of leaking oil please wear gloves providing adequate chemical resistance specifically to aromatic hydrocarbons while cleaning Ventilate the room in case of leaking oil...

Page 15: ...on the frequencies and intensities used these contractions can be adjusted to different strengths The effect is based on the principle that repeated contractions of muscles leads to muscle growth thr...

Page 16: ...Switching on the device Switch on the device using the on off switch 10 Caution Note that connecting the mains cable while the on off switch is on may cause malfunctions Caution If the applicator is u...

Page 17: ...s carried out and the home screen opens Default screen After performing the self test the Z Field Dual will automatically switch to the default screen Selecting the configuration By pressing the confi...

Page 18: ...Volume Adjust the volume using the arrow keys to the left and right D Language Select the language using the arrow keys to the left and right E Administrator Only for service partners F Version Displa...

Page 19: ...tion device operators assistants and patients must not carry any objects on them that may be affected by the magnetic field Note When operating the device do not use objects such as wristwatches mobil...

Page 20: ...gbelt serves to fix the applicators and helps to minimize the movements of the applicators during the treatment The applicators should be placed into the pocket of the fixingbelt The underside of the...

Page 21: ...on changes to STOP Pressing this button again will stop the treatment and reset the treatment time as well as the applied energy M Channel selection Selection of channel 1 and or channel 2 Channels ca...

Page 22: ...manual mode is selected this screen displays a graph representing the selected parameters Manual mode The manual mode has five freely storable modes the user can set the parameter values directly The...

Page 23: ...een appears the user can set the parameter values After pressing Name for about 2 seconds an on screen keyboard appears and the pre programmed name can be changed Pressing the Parameter section will d...

Page 24: ...fferent parameter sets The manual mode works sequentially from sub mode 1 up Pressing the pen icon at the end of the parameter setting screen at the top right adds a sub mode To remove a sub mode sele...

Page 25: ...Channel 2 0 5 1 0 T 20 Modes Programme mode A1 A20 Region mode M1 M20 Manual Mode U1 U20 Transport Transport the device in an upright position 1 unit of packaging 1 device Environmental conditions fo...

Page 26: ...ve rough soiling to prevent it from drying Needed aids Washing machine for chemical disinfection Laundry net Disinfectant detergent which is compatible to the components of the fixingbelt see above an...

Page 27: ...tergent according to the manufacturer s instructions but is not dripping wet until the dirt is removed Disinfection we recommend disinfecting at a minimum after each patient and at the end of the day...

Page 28: ...The product Z Field Dual MFG 03 bears the CE marking 0123 in accordance with EC Directive 93 42 EEC concerning medical devices Manufacturer Zimmer MedizinSysteme GmbH Junkersstra e 9 89231 Neu Ulm Ge...

Page 29: ...S 50 70 cm 54209410 1 Fixingbelt M 80 105 cm Accessories Catalogue number 10102932 1 instructions for use 54209400 1 Fixingbelt S 50 70 cm 54209410 1 Fixingbelt M 80 105 cm Subject to technical chang...

Page 30: ...ted in accordance with the instructions for use enhancements readjustments or changes are only made by persons authorised by Zimmer MedizinSysteme GmbH before using the device the user has satisfied h...

Page 31: ...y a person authorised by Zimmer MedizinSysteme GmbH To ensure safe use have the system including its internal components and output voltage checked once a year by a person authorised by the company Pl...

Page 32: ...ures which are necessary as a result The device is not listed in attachment 2 of the MPBetreibV In Germany the German Social Accident Insurance DGUV Regulation 3 Electrical systems and equipment as am...

Page 33: ...zontal position If the applicator overheats operation will be temporarily suspended and the Overheating message will be displayed see 23 If this message appears remove the applicator from the patient...

Page 34: ...that the output level is set by pressing the start button and then turning the knob Chapter 6 Operating instructions The message Overheating will be displayed When the applicator is in an upright pos...

Page 35: ...ice can be reached via the head office in Neu Ulm Head office Zimmer MedizinSysteme GmbH Junkersstra e 9 89231 Neu Ulm Germany Tel 49 731 9761 291 Fax 49 731 9761 299 export zimmer de www zimmer de De...

Page 36: ...nt could result in increased electromagnetic emission or decreased electromagnetic immunity of this equipment and result in improper operation The device MFG 03 contains no interchangeable components...

Page 37: ...3 2 Class A Emissions of voltage fluctuations flickers in accordance with IEC 61000 3 3 Not performed Table 2 Guidance and Manufacturing Declaration Electromagnetic Immunity The device MFG 03 is inte...

Page 38: ...phase at 0 Voltage interruptions accordance with IEC 61000 4 11 0 UT 250 300 cycles 0 UT 250 cycles Magnetic field of supply frequency 50 60 Hz in accordance with IEC 61000 4 8 30 A m 50 Hz oder 60 H...

Page 39: ...430 470 GMRS 460 FRS 460 Pulse Modulation 18 Hz 2 3 28 710 704 787 LTE Band 13 17 Pulse Modulation 217Hz 0 2 3 9 745 780 810 800 960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse Modulation...

Page 40: ...Material Safety Data Sheet 16 Page 30...

Page 41: ...Material Safety Data Sheet 16 Page 31...

Page 42: ...Material Safety Data Sheet 16 Page 32...

Page 43: ...Material Safety Data Sheet 16 Page 33...

Page 44: ...Material Safety Data Sheet 16 Page 34...

Page 45: ...Material Safety Data Sheet 16 Page 35...

Page 46: ...Material Safety Data Sheet 16 Page 36...

Page 47: ...eme GmbH Junkersstra e 9 89231 Neu Ulm Germany Tel 49 7 31 97 61 291 Fax 49 7 31 97 61 299 export zimmer de www zimmer aesthetics com Instructions for use EN 10 102 932 VK 0819 I MFG 03 I Version 2 I...