Function Test
15
Page 28
Thermo
Pro
performs a self-test after it is switched on, which checks all the
internal components.
An error message will appear when a fault occurs.
An extended function test as described below can also be performed.
This test should be performed monthly or if there are any doubts about the
functionality of the device.
Function test
Select continuous operating mode and set an output power of 5 W.
Move your forearm into the treatment field and make sure that the coupling
indicator displays at least 50%.