Zimmer MIS Multi-Reference 4-in-1 Femoral Instrumentation Surgical Technique
34
Option 1
Tibial Position based on
Anatomic Landmarks
The position of the tibial component
can also be determined based on
anatomic landmarks prior to trial
reduction. Select the appropriate
Pegged or Stemmed Tibial Sizing
Plate/Provisional that provides the
desired tibial coverage (Fig. 8c). Please
refer to the
Zimmer
®
NexGen
®
MIS
Tibial Component Cemented Surgical
Technique
(97-5950-002-00) for
complete product information and
instructions for the MIS Tibial
Stemmed component.
Fig. 8c
Pegged Tibial Sizing Plate
Stemmed Tibial Sizing Plate
Attach the Universal Handle to the
selected Tibial Sizing Plate/Provisional
by depressing the button on the handle
and engaging the dovetail on the
handle with the dovetail on the sizing
plate/provisional and secure it by
tightening the thumbscrew (Fig. 8d).
Fig. 8d
Generally, the handle aligns with the
anterior aspect of the tibia. Rotate the
sizing plate/provisional so the handle
points at, or slightly medial to, the
tibial tubercle (Fig. 8e). The Alignment
Rod can be used to aid in double
checking varus/valgus alignment.
Pin the plate in place with two Short
Head Holding Pins.
Fig. 8e