4.3.1 Calibration accessories
CAUTION
×
The calibration plate is a certified measuring tool; hold by the base when fitting and removing.
×
Never touch the ceramic plate (test marks on the calibration plate could be damaged).
×
Do not drop the calibration plate (risk of breakage)
×
Never wipe the ceramic plate or clean it with an abrasive substance
(scratches on the plate may significantly impair the measurement result)
4.3.2 Preparation
×
Lay out the calibration plate, mounting plate and
4 spacer plates.
×
Start by clicking the
“Calibration”
button
×
Select the connected scanner
1
×
Confirm the selection by clicking “OK”
2
NOTE
The calibration procedure repeats itself 5 times at different heights.
4.3.3 Steps 1 to 5
Calibration plate
white
4 × Spacer plates
1 × mounting plate
1
st
step
×
Insert the calibration
plate with the base
plate
(#1)
in the
scanner.
×
Confirm selection by
clicking
“OK”
1
×
Insert first spacer
plate
(#2)
×
Confirm selection by
clicking
“OK”
1
×
Insert first spacer
plate
(#3)
×
Confirm selection by
clicking
“OK”
1
×
Insert first spacer
plate
(#4)
×
Confirm selection by
clicking
“OK”
1
×
Insert first spacer
plate
(#5)
×
Confirm selection by
clicking
“OK”
1
2
nd
step
3
rd
step
4
th
step
5
th
step
1
1
1
1
1
1
2