22
Reference conditions
Ambient temperature :
+ 23
0
C + 2 K
Relative humidity :
45% ... 55 % RH
Frequency of measured
quantity
45Hz ...65 Hz
Waveform of the measured
quantity
sinusoidal
Battery voltage
8 V + 0.1 V
Measuring voltage for resistance measurement
Voltage U
x
across the resistance R
x
to be measured as a
function of measuring range and display.
Measuring current I
x
as a function of the displayed voltage
U
x
across the device under test
1,0
0,8
0,6
0,4
0,2
0
0
1000
2000
3000
Digit / digits
U
x
V
30 MW
3 MW
300 kW
30 kW
300 W
3 kW
30 W
2,0
1,5
1,0
0,5
0
0,5
1
1,5
2
I
x
mA
U
x
V
Measuring current for diode test or continuity test
7
The MIN/MAX function is switched OFF, when the MIN/MAX pushbutton
(3) is pressed for approximately 1s, or when the function selector switch
(6) is operated, or when the meter is turned OFF and ON again.
Notes :
The 30 mV
and 300 mV
measuring ranges can only be selected
manually with the ’’ AUTO/MAN ’’ pushbutton (4) !
On the 1000 V r ange, an intermittent sound signal warns you, when the
measured value exceeds the upper range limit.
Caution :
Ensure current measuring range (’’mA’’ or ’’A’’) is not selected for voltage
measurement ! When the cut-out rating of the fuses is exceeded because
of incorrect operation, a dangerous situation exists!.
Zero adjustment on the 30 mV
measuring range
Connect the test leads to the meter and join the free ends.
After having selected the measuring range, briey press the yellow multi-
function pushbutton (5).
The meter acknowledges zero setting by a sound signal, the LCD shows
’’00.00’’ (+ 1 digit) and the decimal point ashes. The displayed voltage
at the instant the pushbutton is pressed, is used as reference value
(max + 200 digits) it is automatically deducted from the values measured
thereafter.
The zero adjustment is cleared when ;
By pressing the yellow multifunctionpushbutton (5) for a long time,
clearance is acknowledged by the two sound signal.
By switching the instrument OFF.
8. Voltage measurement
According to the voltage to be measured , set the function selector
switch (6) to V
~,
V
or V
.
Connect the test leads as shown. The ’’
’’ socket should be
connected to the lowest potential ground available.
Function
MIN/MAX
2 x Short,
30 mV/
300 mV
and
0
C
1 x short
short
short
Short
Long
DATA
MIN/MAX
(3)
Measured
Values
MINandMAX
Stored
Storage Continued
in the background,
new MIN / MAX.
values are displayed
Same as 1.,
Stored Values are
not cleared
Cleared
actual
meas-
ured
value
stored
MIN
value
stored
MAX
value
same
as 1.
Cleared
MIN
and
MAX
ash
MIN
MAX
same
as 1.
Cleared
Meteracknowledgement
Display
MIN
MAX
Meas-
uring
ranges
Sound
Signal
1.
Activate and
Store
2.
Store and
display
3.
Return to 1.
Reset
1 x
1 x
1 x
1 x
2 x
Meas.
Value
digital
V
A
W
F,Hz, %
0
C
V
A
W
F,Hz,%
0
C
Same
as 1.