![Zhermack SAB 1000 G User Manual Download Page 26](http://html1.mh-extra.com/html/zhermack/sab-1000-g/sab-1000-g_user-manual_3635676026.webp)
133 of 180
MAINTENANCE AND USE MANUAL
SQM25 N - SQM25 SN - SAB1000G
EN
CHAPTER 8: MAINTENANCE
8.1
CLEANING
The machine exterior can normally be kept clean just by wiping it with a dry cloth. If necessary,
wet the cloth just a little water or use a non-degreasing detergent.
The operator must ensure that the machine is kept clean and free from any foreign bodies such
as dirt, water, non-conducting liquids etc.
The machine must be cleaned at the end of every work session to remove all residues.
Cleaning must be done with the machine at standstill and positioned in a stable manner.
Never use inflammable, corrosive, alkaline or toxic liquids when cleaning the
machine.
Eye goggles and a face mask must be worn when using compressed air for
cleaning purposes. Do not allow any persons near the machine as they could
be hit by flying dust.
Before cleaning the machine exterior always turn the main switch to OFF. It is
strictly forbidden to clean the machine while it is connected to the mains power
supply.
If the model trimmer is to remain idle for a prolonged period it is advisable to disconnect it so as
to prevent the motor overheating.
8.1.1
Replacing the nylon filter bag on the
SAB 1000 G
To change the nylon bag filter in the
SAB 1000 G
aspirator (usually done weekly) proceed as
follows:
• switch off the aspirator via the main switch and disconnect it from the trimmer.
• undo the catches (
H
) to open the aspirator and remove the motor block (
A
).
• remove the felt filter (
L
) and detach the suction hose (
F
) from the connector.
• remove the filter (
I
).
When changing the filter make sure no powder gets inside the aspirator; if it does, clean
thoroughly.
• insert new filter.
• reposition the felt motor protection filter (
L
) so that its edges rest on the aspirator body (
E
).
• place the motor block (
A
) on top of the aspirator body (
E
) and make sure the felt filter is
positioned properly in the motor block.
• close the aspirator by locking the catches (
H
) back in place.
• re-attach the suction hose (
F
) to the connector.
• insert the plug (
G
) in a mains socket, then switch on via the main switch (
B
).