Power supply

Battery powered mode:

6VDC 4×AA-size batteries

AC adaptor powered mode:

100-240 V~, 50-60 Hz, 400 mA
(Not Included) 

Display mode

Blue LCD with white backlight  

V.A. = 78mm x 92mm

Measurement mode

Oscillographic testing mode

Measurement range

Pressure: 0-40kpa(0~300mmHg)
pulse value:(40-199)times/minute

Accuracy

Pressure:

5

-40

 within ±0.4 kPa (3 mm Hg)

Pulse Value: ±5%

Working condition

Temperature:5

-40

Relative Humidity ≤85%RH

Atmospheric Pressure: 86-106 kPa

Storage & transportation

condition

Measurement perimeter

of the upper arm

About 22cm - 32cm

Weight

Approx.499g(Excluding the dry cells)

External dimensions

Attachment

Approx.120×160×69mm

4×AA batteries,user manual, storage bag  

Mode of operation

Continuous operation

Degree of protection

Type BF applied part

IP Classification

IP22

WARNING: No modification of this equipment is allowed.

Temperature:-20

-60

Relative Humidity 10%-93%RH

Atmospheric Pressure: 50-106 kPa

Software Version

V01

Device Classification

Battery Powered Mode: Internally Powered ME Equipment

AC Adaptor Powered Mode: Class II ME Equipment

Authorized Components

Please use the ZEWA 

authorized adaptor

(Not Included)

2

.Storage bag.

Complied Standards List

Contact Information

UE08WCP-060100SPA

100-240V

50-60Hz,400mA

6V      1A

Adaptor
Type:

Input:

Output:

(Expected Service Life: 50,000 Hours)

For more information about our products, please visit www.zewa.com.you can get 

customer service, usual problems and customer download, Zewa will serve you 

anytime.

 

AUTHORIZED COMPONENTS

SPECIFICATIONS

 

 

ISO 14971:2007
EN 980:2008
EN 1041:2008

Risk Management
Labeling
User Manual

Generl Requirements

for Safety

Performance and

Clinical 

Requirements

Electromagnetic

Compatibility

IEC 60601-1:2005/A1:2012

ANSI/AAMI/IEC 80601-2-30:2009

ANSI/AAMI/ISO 81060-2:2009

IEC 60601-1-2:2007

Manufactured for Zewa, Inc.

12960 Commerce Lakes Drive#29 

Fort Myers, FL 33913 USA

www.zewa.com

Toll Free Customer Service:

1-888-993-3592

[email protected] 

Summary of Contents for UAM-910BT

Page 1: ...and keep the manual well for further reference in case you have problems Thank you very much for selecting Zewa Blood Pressure Monitor UAM 910BT Arm Type Version 1 0 Contains FCC ID OU9AW8001 LS Manu...

Page 2: ...ood pressure varies even in one day Why the blood pressure I get from the hospital is different from home TROUBLESHOOTING 19 SPECIFICATIONS 20 AUTHORIZED COMPONENTS 21 COMPLIED STANDARDS LIST 21 EMC G...

Page 3: ...ications for Use Contraindications Measurement Principle This product uses the Oscillometric Measuring method to detect blood pressure Before every measurement the unit establishes a zero pressure equ...

Page 4: ...stop inflation The equipment is not AP APG equipment and not suitable for use in the presence of a flammable anesthetic mixture with air of with oxygen or nitrous oxide The operator shall not touch o...

Page 5: ...der to achieve the best performance and protect you monitor please use the authorized specified battery and power adaptor CAUTION 1 Open the battery door 2 Insert the batteries according to the polari...

Page 6: ...1 button again to confirm HOUR Then the numeral representing MINUTE blinks Please proceed to time setting before your initial use so as to ensure each piece of record are labled with a time stamp The...

Page 7: ...side of arm in line with the little finger 2 The cuff should be sung but not too tight You should be able to insert one finger between the cuff and your arm 3 Sit comfortably with your left arm resti...

Page 8: ...tor is off press the User 1 button to turn on the monitor 2 Press the User 1 button to power off otherwise it will turn off with one minute A When finish the whole measurement press another user butto...

Page 9: ...clearing the memory you may press the other button rather than Query button to turn off the monitor or wait until the monitor shuts off INFORMATION FOR USER DATA MANAGEMENT Data Transmission 1 Once th...

Page 10: ...to 15 this equipment will light up the IHB symbol on the screen when displaying the measuring result CAUTION The appearance of the IHB icon indicates that a pulse irregularity consistent with an irre...

Page 11: ...ly asked questions for problems you may encounter with your blood pressure monitor If the product isnot operating as you think it should check here before fit servicing PROBLEM SYMPTOM CHECK THIS REME...

Page 12: ...ve Humidity 10 93 RH Atmospheric Pressure 50 106 kPa Software Version V01 Device Classification Battery Powered Mode Internally Powered ME Equipment AC Adaptor Powered Mode Class II ME Equipment Autho...

Page 13: ...he user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving antenna Increase the separation between the product and the receive...

Reviews: