background image

11

10

Pair-up the Blood Pressure Monitor 
with Your Bluetooth Device

1

.Turn on Bluetooth and the Zewa app. Make sure both are ON when 

pair-up is proceeding.

2

.When the monitor is OFF, press 

start pair-up. The symbol         

and the symbol         will be 

shown on the LCD alternatively, 

indicating pair-up is proceeding.

 

 

If SUCCEED

, symbol         

will be shown on the LCD.

 

If FAIL

, symbol         will 

be shown on the LCD.

 

3

.The monitor will shut off after Pair-up 

process is complete. 

Bluetooth Module No.: 

AW8001

RF Frequency Range:

 2402 MHz to 2480 MHz

Output Power Range: 

0 dBm

Supply Voltage: 

1.8-3.6 V

Transmitting Distance: 

10 meters

and hold the     button (User 2) to 

Resting for 5 minutes before 

measuring.

Wait at least 3 minutes between 

measurements. This allows your 

blood circulation to recover.

For a meaningful comparison, 

try to measure under similar 

conditions. For example, take daily 

measurements at approximately 

the same time, on the same arm, 

or as directed by a physician.

Tie the Cuff

1

.Tie the cuff on your upper arm, 

the position the tube off-center 

toward the inner side of arm in 

line with the little finger.

2

.The cuff should be sung but not 

too tight. You should be able to 

insert one finger between the 

cuff and your arm.

3

.Sit comfortably with your left 

arm resting on a flat surface. 

2~3cm

MEASUREMENT

BEFORE YOU START

CAUTION

Interference may occur in the vicinity of equipment marked with the following symbol

      . And UAM-910BT may interfering vicinity electrical equipment.
Sensitive people, including pregnant women pre-eclamptic and those who implanted 

medical electronic instruments, should avoid using the unit whenever possible.
Keep the monitor at least 20 centimeters away from the human body (especially the 

head) when the data transmission is proceeding after measurement.
To enable the data transmission function, this product should be paired to Bluetooth 

end at 2.4 GHz.

How to mitigate  possible interference?

1. The range between the device and BT end should be reasonably close, from 1 

meter to 10 meters. Please ensure no obstacles between the device and BT end

so as to obtain quality connection and to lower the RF output range.

2. To avoid interference, other electronic devices (particularly those with wireless 

transmission / Transmitter) should be kept at least 1 meter away from the monitor.

Summary of Contents for UAM-910BT

Page 1: ...and keep the manual well for further reference in case you have problems Thank you very much for selecting Zewa Blood Pressure Monitor UAM 910BT Arm Type Version 1 0 Contains FCC ID OU9AW8001 LS Manu...

Page 2: ...ood pressure varies even in one day Why the blood pressure I get from the hospital is different from home TROUBLESHOOTING 19 SPECIFICATIONS 20 AUTHORIZED COMPONENTS 21 COMPLIED STANDARDS LIST 21 EMC G...

Page 3: ...ications for Use Contraindications Measurement Principle This product uses the Oscillometric Measuring method to detect blood pressure Before every measurement the unit establishes a zero pressure equ...

Page 4: ...stop inflation The equipment is not AP APG equipment and not suitable for use in the presence of a flammable anesthetic mixture with air of with oxygen or nitrous oxide The operator shall not touch o...

Page 5: ...der to achieve the best performance and protect you monitor please use the authorized specified battery and power adaptor CAUTION 1 Open the battery door 2 Insert the batteries according to the polari...

Page 6: ...1 button again to confirm HOUR Then the numeral representing MINUTE blinks Please proceed to time setting before your initial use so as to ensure each piece of record are labled with a time stamp The...

Page 7: ...side of arm in line with the little finger 2 The cuff should be sung but not too tight You should be able to insert one finger between the cuff and your arm 3 Sit comfortably with your left arm resti...

Page 8: ...tor is off press the User 1 button to turn on the monitor 2 Press the User 1 button to power off otherwise it will turn off with one minute A When finish the whole measurement press another user butto...

Page 9: ...clearing the memory you may press the other button rather than Query button to turn off the monitor or wait until the monitor shuts off INFORMATION FOR USER DATA MANAGEMENT Data Transmission 1 Once th...

Page 10: ...to 15 this equipment will light up the IHB symbol on the screen when displaying the measuring result CAUTION The appearance of the IHB icon indicates that a pulse irregularity consistent with an irre...

Page 11: ...ly asked questions for problems you may encounter with your blood pressure monitor If the product isnot operating as you think it should check here before fit servicing PROBLEM SYMPTOM CHECK THIS REME...

Page 12: ...ve Humidity 10 93 RH Atmospheric Pressure 50 106 kPa Software Version V01 Device Classification Battery Powered Mode Internally Powered ME Equipment AC Adaptor Powered Mode Class II ME Equipment Autho...

Page 13: ...he user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving antenna Increase the separation between the product and the receive...

Reviews: