This section includes a list of error messages and frequently
asked questions for problems you may encounter with your blood
pressure monitor. If the product is not operating as you think it
should, check here before arranging for servicing.
Digital LCD V.A.80mm*60mm
Approx.250g(Excluding the batteries)
Approx.102mm*143mm*73mm
6VDC 4×AA batteries
A01
4×AA batteries,user manual ,Carry bag
6V 1A (Not inicluded)
Battery powered mode:
AC adaptor powered mode:
(Please only use the recommended AC
adaptor model).
About 22cm~42cm
Type BF applied part
WARNING: No modification of this equipment is allowed.
Power supply
Display mode
Measurement mode
Oscillographic testing mode
Measurement range
Measurement perimeter
of the upper arm
Weight
External dimensions
Attachment
Mode of operation
Continuous operation
Degree of protection
Protection against
ingress of water
Accuracy
Normal working condition
Storage & transportation
condition
Software Version
Pressure:
5°C-40°C
within±3mmHg(0.4kPa)
Pulse value: ±5%
Rated cuff pressure:
0mmHg~299mmHg(0kPa ~ 39.9kPa)
Measurement pressure:
SYS: 60mmHg~230mmHg (8.0kPa~30.7kPa)
DIA: 40mmHg~130mmHg (5.3kPa~17.3kPa)
Pulse value: (40-199)beat/minute
IP21 It means the device could protected against
solid foreign objects of 12.5mm and greater, and
protect against vertically falling water drops.
Device Classification
Battery Powered Mode:
Internally Powered ME Equipment
AC Adaptor Powered Mode: Class II ME Equipment
A temperature range of : +5°C to +40°C
A relative humidity range of 15% to 90%,
non-condensing, but not requiring a water vapour
partial pressure greater than 50 hPa
An atmospheric pressure range of :
700 hPa to 1060 hPa
Temperature:-20°C to +60°C
A relative humidity range of ≤ 93%, non-condensing,
at a water vapour pressure up to 50hPa
PROBLEM SYMPTOM CHECK THIS
REMEDY
No power
Low
batteries
Error
message
Display
will not light up.
Batteries are exhausted. Replace with new batteries
Insert the batteries
correctly
Replace with new batteries
Batteries are inserted
incorrectly.
Batteries are low.
E 1 shows
E 3 shows
E20 shows
EExx,shows on
the display.
A calibration error
occurred.
Retake the measurement.
If the problem persists,
contact the retailer or our
customer service
department for further
assistance.Refer to the
warranty for contact
information and return
instructions.
AC adaptor is inserted
incorrectly.
Insert the AC adaptor
tightly
Display is dim or
shows
Warning
message
“out ” shows
Out of measurement
range
the measurement result
is out of the measurement
range (SYS:60mmHg to
230mmHg; or DIA:40mmHg
to 130mmHg;or Pulse:
40-199 pulse/minute)
The cuff is not secure.
Refasten the cuff and then
measure again.
Relax for a
moment and then
measure again.
The pressure of the
cuff is excess.
E 10 or
E 11 shows
movement can affect the
measurement.Relax for a
moment and then measure
again.
The measurement
process does not detect
the pulse signal.
Loosen the clothing on the
arm and then measure
again.
Measure incorrectly.
The monitor detected
motion,talking,or the
pulse is too poor while
measuring.
E21 shows
Relax for a moment and
then measure again.
25
SPECIFICATIONS
TROUBLESHOOTING
24