COOLTONE User Manual
Appendix B
:
System Specifications
-C
45
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic
Environment - Guidance
Voltage dips in
accordance with
IEC 61000-4-11
0 % U
T;
0,5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % U
T;
0,5 cycle
At 0°, 45°, 90°, 135°,
180°,
225°, 270° and 315°
The supply voltage quality must
correspond to that of a typical
commercial or hospital
environment. If the user of the
device CoolTone requires
continued operation, even in the
case of interruptions in the
power supply, it is recommended
that the device CoolTone be
powered from an uninterrupted
power supply or a battery.
0 % U
T;
1 cycle and 70%
U
T;
25/30 cycles
Single phase: at 0°
0 % U
T;
1 cycle and 70%
U
T;
25/30 cycles
Single phase: at 0°
Voltage
interruptions
accordance with
IEC 61000-4-11
0% U
T;
250/300 cycle
0% U
T;
250/300 cycle
Magnetic field of
supply frequency
(50/60 Hz) in
accordance with
IEC 61000-4-8
30 A/m
60 Hz
30 A/m
60 Hz
Magnetic fields at mains
frequency should have the
typical values found in a
business or hospital
environment.
Note: U
T
is the mains AC
Voltage before application of the
test level
Conducted
Disturbances
induced by RF
fields according
IEC 610004-6
3 V
0,15 MHz to 80
MHz
6 V in ISM Band between
0,15 MHz and 80 MHz
80% AM at 1 kHz
3 V
0,15 MHz to 80 MHz
6 V in ISM Band
between 0,15
MHz and 80 MHz
80% AM at 1 kHz
In the vicinity of devices, bearing
the following symbol,
interference is possible:
Radiated RF EM
fields according
IEC 610004-3
3 V/m
80 MHz-2,7 GHz
80% AM to 1 kHz
3 V/m
80 MHz-2,7 GHz
80% AM to 1 kHz