Cleaning instructions
The ZEISS VISUSCOUT 100 is a precision optics instrument that should be
handled with care. Please note following cleaning instructions:
•
Shut down device before cleaning it
•
Remove cradle from mains before cleaning it
•
Disinfect housing with soft cloth moistened with alcohol (e.g. 70%
ethanol). Avoid touching System connectors in the handset and cradle.
•
Lenses may be cleaned with cleaning cloth. Also, moist-cleaning tissue
such as Hama Pro-Optic® cleaning tissue can be used.
The VISUSCOUT 100 is not intended to be sterilized.
CAUTION
Do NOT use the following chemicals for cleaning:
•
mineral spirits, paint thinner, benzene, gasoline, lamp oil
•
Strong / corrosive acids (such as sulphuric acid)
•
Strong / corrosive bases (such as sodium hydroxide)
•
Bleaching agents
•
Nail polish remover
CAUTION
Clean the eye cup before each use on a new patient to avoid contamination.
Clean the eye cup before each use on a new patient:
•
disinfect the eye cup with soft cloth moistened with alcohol (e.g. 70%
ethanol), or
•
soak the eye cup in glutaraldehyde-based solution, or hydrogen peroxide
and peracetic acid solution
•
rinse the eye cup under running water
•
dry the eye cup (e.g. with clean paper towel) before subsequent use
If replacement for the eye cup is needed, please contact ZEISS customer service
or your own retailer.
Eye cup should be replaced if/when:
•
the eye cup is discolored
•
the eye cup form is deteriorated
•
the eye cup is shattered, cracked or disintegrated