000000-1363-329 SL 120 11.07.2006
30
Slit Lamp SL120
Line voltage
100 V -10 % ... 240 V +10 %, 50 / 60 Hz
U.S. and Canada only: single-phase connection
120 V AC with plug type NEMA 5-15P
Power consumption
75 VA
Protection class
I
Instrument type
B (acc. to EN 60601-1)
Enclosure protection
IP 20
Environment conditions
for intended use
Temperature:
+10 °C ...+35 °C
Humidity:
30 ... 75 % (no condensation)
Air pressure:
800 ... 1060 hPa
Storage conditions
(in original packing)
Temperature:
-10 °C ...+55 °C
Humidity:
10 ... 95 %
Air pressure:
700 ... 1060 hPa
Shipment conditions
(in original packing)
Temperature:
-40 °C ...+70 °C
Humidity:
10 ... 95 %
Air pressure:
500 ... 1060 hPa
Mode of operation
continuous operation
Light source for
projection
6 V / 20 W, halogen lamp
Lamp voltage
continuously adjustable
Fixation lamp
5 V / 1.5 W
Magnification
5x, 8x, 12x, 20x, 32x (with high eyepoints 10x)
6x, 10x,16x, 25x, 40x (with high eyepoints 12.5x)
Field of view
40 mm ... 6 mm (with high eyepoints 10x)
31 mm ... 5 mm (with high eyepoints 12.5x)
Eyepiece magnification
either 10x or 12.5x high eyepoint,
±8 D compensation of ametropia
Slit length
0 to 14 mm continuously progressive
Slit width
in steps at 0.3 / 3.5 / 8 / 14 mm; 1 ... 6 mm
continously
Slit rotation
±90° continously
Slit decentration
±4° horizontal, click-stop at 0°
Swivel range of slit
prism
180°, scale for angular difference
click-stops at -10°, 0°, +10°
Angle of incidence
0° or 0° ... 20° with tild prism head
Filters
blue, green (red-free); heat absorbing filter (fixed);
swing-in diffusor in front of the prism
Free working distance
65 mm
Travel of instrument
base
30 mm (vertical), 110 mm (X), 90 mm (Y)
Vertical travel of chin
rest
58 mm
Mass
9.75 kg (basic unit), 1.25 kg (headrest)
Specification
Summary of Contents for SL 120
Page 2: ......
Page 4: ...000000 1363 329 SL 120 11 07 2006 2...
Page 36: ...000000 1363 329 SL 120 11 07 2006 2...
Page 68: ...000000 1363 329 SL 120 11 07 2006 2...
Page 100: ...000000 1363 329 SL 120 11 07 2006 2...
Page 131: ......