Start-Up and Operation
Primovert
16
415510-7244-001
12/2014
3.1.3
Inserting the Condenser
•
Remove the condenser cap.
•
Insert the condenser (Fig. 4/
4
) from the front
into the guide (Fig. 4/
3
) and push it in to stop.
•
Lock the condenser by tightening the socket
head cap screw or the knurled thumb screw
(Fig. 4/
1
), which is equally supplied with the
microscope.
Recommendation:
If condensers are frequently exchanged
or whenever larger glass vessels are to
be used frequently without a
condenser, the knurled thumb screw
supplied rather than the socket head
cap screw (Fig. 4/
2
) should be
employed to lock the condenser.
3.1.4
Inserting the Filter Slider
•
Put the color filter (Fig. 5/
1
) or the green
interference filter (diameter d=45 mm) into the
opening of the filter slider.
As the filters are clamped in the filter
slider only slightly, not twist-proof, the
filter slider should not be turned or
tilted in order to avoid that the filters
fall out and get damaged.
•
Push the 2-position filter slider (Fig. 5/
2
) from
the left or the right side into the slot (Fig. 5/
3
)
of the upper part of the stand until it engages
in the desired position.
When using the Primovert iLED in
fluorescent mode, use the yellow filter
inserted into the filter slider (included
in the delivery). It prevents scattered
fluorescent background generated by
unwanted transmitted LED source
phosphor.
Fig. 4
Inserting the condenser
Fig. 5
Inserting the filter slider