Instructions for Use
6 Before every use
OPMI LUMERA 700
6.1 Safety during preparation
G-30-1673-en - 14.0 - 2018-07-12
123 / 236
6 Before every use
6.1 Safety during preparation
WARNING!
Risk of injury caused by attached fundus viewing system!
Incorrect handling of a fundus viewing system mounted on the
underside of the microscope, or activation of the fast focus, may
cause injury to the patient’s eye.
u
Before using a fundus viewing system, make sure that the
space available for movement is greater than the downward
travel path of the microscope.
u
Prior to using the fundus viewing system, reset the focus.
[
CAUTION!
Risk of injury caused by incorrect software settings!
Incorrect software settings may lead to unexpected behavior of the
device and may cause injuries to the patient.
u
Check the selected user before each use.
u
Only make changes to your own user profile.
NOTE
Loss of functionality caused by hardware or software faults!
u
Arrange suitable precautions, irrespective of the application, to
be able to end the operation or application without use of this
microscope.
u
Keep a backup lamp or an OR illuminator ready at hand.
NOTE
Light source failure due to overheating!
If the ventilation openings are covered, the device can accumulate
heat, and the light source may stop working.
u
Never cover any ventilation openings.
Summary of Contents for opmi lumera 700
Page 1: ...OPMI LUMERA 700 Software Release 3 1 Instructions for Use...
Page 10: ...Empty page for your notes...
Page 14: ...Empty page for your notes...
Page 30: ...Empty page for your notes...
Page 130: ...Empty page for your notes...
Page 200: ...Empty page for your notes...
Page 222: ...Empty page for your notes...
Page 226: ...Empty page for your notes...
Page 230: ...Empty page for your notes...
Page 235: ...Empty page for your notes...