OPMI® 1FR XY Surgical Microscope on S7 Floor Stand
Issue 4.0
Printed on 14. 11. 2007
Control panel
Note:
The OPMI 1FR XY on the S7 floor stand is not provided with a motorized
zoom function.
1
LED display
Display of the selected settings for lamp brightness and the motor
speed of the focusing and XY-coupling functions.
2
Four LEDs
for assigning the LED display to the function selected via arrow keys
(4): lamp brightness, motor speed for focusing and X-Y coupling. Only
one diode is lit at a time.
3
Plus / minus keys
for increasing or decreasing the value of the function selected using
keys (4): lamp brightness, motor speed for focusing and X-Y coupling.
The following setting options are available:
Lamp brightness
Adjustment range: 0.1...1.5
Level 1.0 corresponds to the spectral density of L
B
= 500 mW / cm² sr
as specified in the ISO 10936-2 standard draft for surgical micros-
copes used in ophthalmic applications.
Motor speed
Adjustment range: 1...10
Level 1 corresponds to the lowest, level 10 to the highest motor speed.
4
Arrow keys
for selecting one of the following functions: lamp brightness, motor
speed of focusing and X-Y coupling. The relevant LED is lit.
Notes:
The value set is saved after 10 seconds and displayed the next time
the unit is turned on.
If no entry is made via the arrow or plus/minus keys, the unit will switch
to the lamp brightness display after 15 seconds at the latest.
5
Power switch
This switch is used to switch the unit on and off. When the unit has
been switched on, the green lamp in the switch is lit.
The power switch also contains an automatic circuit breaker. In the
event of a short inside the unit, the circuit breaker is activated and
switches off the unit.