Safety Measures
KSK 150 FC
Version 7.1
Page 10
G-30-1158-en
• Do not use power-operated devices included in the delivery package
– in explosive atmospheres,
– at a distance of less than 25 cm from flammable anesthetics or volatile
solvents such as alcohol, benzine or similar substances.
• Do not use or store the system in damp rooms. Do not expose the system
to water splashes, dripping water or sprayed water.
• Ensure that fluids cannot penetrate into the device.
• Never directly look with binocular tube, object lens or an eyepiece into the
sun.
A potential equalization connection can be found on the underside of the car-
rier arm (see IEC 60601-1:2005 + A1:2012). This makes it possible to connect
other active devices to the same earth potential or it can be used for the re-
dundant connection with the grounding conductor.
• Always switch off the system before connecting it to / disconnecting it
from line power, for cleaning its surface or if it will not be used for a pro-
longed period of time.
• Please note the information on electromagnetic compatibility (EMC) in the
chapter on "System data" and the section on "EMC - electromagnetic
compatibility".
• The maximum brightness of the lamp(s) can drop over time (normal device
attribute).
• Transport the device over long distances (e.g. relocation, return for repair)
only in its original packaging or special return packaging. Please contact
your dealer or Carl Zeiss Service.
• Pay attention to the places on the device identified by the information la-
bel "Risk of crushing". In these places there a risk of getting your fingers
crushed! Do not touch these areas while the device is being moved.
• If a failure occurs which you cannot correct with the aid of the chapter
"What to do in the event of malfunctions"“, attach a sign to the system
stating it is out of order and contact the Carl Zeiss service staff.
• As the operator, have regular safety checks performed on this device to
prevent a reduction in the safety of the device as a result of aging, wear,
etc. See the chapter on "Care and maintenance" and the section on "Safe-
ty check".
Summary of Contents for KSK 150 FC
Page 1: ...ZEISS Colposcope KSK 150 FC Instructions for Use G 30 1158 en Version 7 1 2015 06 11 ...
Page 4: ...KSK150FC Version 7 1 Page 4 G 30 1158 en ...
Page 15: ...Version 7 1 G 30 1158 en Page 15 KSK 150 FC Safety Measures ...
Page 17: ...Version 7 1 G 30 1158 en Page 17 KSK 150 FC Safety Measures 1 5 1 1 1 6 2 2 4 2 3 ...
Page 19: ...Version 7 1 G 30 1158 en Page 19 KSK 150 FC Safety Measures 7 10 8 11 12 9 8 ...
Page 21: ...Version 7 1 G 30 1158 en Page 21 KSK 150 FC Safety Measures 19 18 17 13 14 15 16 ...
Page 27: ...Version 7 1 G 30 1158 en Page 27 KSK 150 FC System Overview 1 3 2 1 3 4 2 5 6 7 ...
Page 29: ...Version 7 1 G 30 1158 en Page 29 KSK 150 FC System Overview 3 4 5 6 7 1 2 8 9 10 ...
Page 31: ...Version 7 1 G 30 1158 en Page 31 KSK 150 FC System Overview 1 3 6 2 5 4 8 10 11 9 12 7 ...
Page 37: ...Version 7 1 G 30 1158 en Page 37 KSK 150 FC Preparations for use ...
Page 45: ...Version 7 1 G 30 1158 en Page 45 KSK 150 FC Preparations for use 1 2 ...
Page 63: ...Version 7 1 G 30 1158 en Page 63 KSK 150 FC Care and Maintenance 4 1 5 3 2 ...
Page 66: ...Care and Maintenance KSK 150 FC Version 7 1 Page 66 G 30 1158 en ...
Page 84: ...System Data KSK 150 FC Version 7 1 Page 84 G 30 1158 en ...
Page 90: ...Index KSK 150 FC Version 7 1 Page 90 G 30 1158 en ...
Page 91: ...Version 7 1 G 30 1158 en Page 91 KSK 150 FC Blank page for your notes ...