Maintenance
Corona process
Place the clamping ring
3
on top of the halogen lamp.
Screw the three screws
1
back into the lamp holder using a TX 10
screwdriver
2
.
Tighten the clamping ring with the three screws until there is no longer any
gap between the clamping ring and the lamp holder.
10
Reinstalling the lamp holder
Carefully position the lamp holder (see Fig. 23/
1
) on the lamp base – being
careful not to touch the light guide in the process (see Fig. 23/
2
).
Make sure that the black mark on the lamp socket is pointing upwards (see
Fig. 25/
4.1
).
Screw the lamp holder back onto the base with the four screws (see Fig. 22/
2
).
Plug the lamp cable (see Fig. 21/
1
) back onto the lamp socket.
11
Inserting the device base in the device cover
(see Fig. 19 – reverse order)
Carefully set down the device base
2
on the carrying handle
3
and into
the device cover
4
.
Use the four hexagon screws
1
to screw the device base together with the
device cover.
12
Calibrating compensation data
Both halogen lamps must be recalibrated after each lamp change (refer to
section 7.3 “Lamp calibration”).
13
Installing the Corona process
Perform work steps 3 to 5 from section 4.2 “Installing the Corona process”
on page 22.
Perform work steps 1 to 3 from section 4.3 “Connecting system components”
on page 24.