background image

Page 

8

Installation Instructions

.

3. Shipping Crates and Pallets

3.1

Dimensions and Weights

The machine components are delivered on transport pallets with insulated packaging. They are delivered in a
shipping crate only for air cargo and overseas shipments.

Note!

L

W

H

Shipping crate for

Crate dimensions

in mm (max.)

m

3

Max. tare wt. (kg)

Max. gross wt. (kg)

L

W

H

Sea/air 

cargo crate

Truck

pallet

w. sea/air cargo 

crate

with truck

pallet

CONTURA G2

X=700, Y=700, Z=600
X=700, Y=1000, Z=600
X=1000, Y=1200, Z=600
X=1000, Y=1600, Z=600
X=1000, Y=2100, Z=600

2820
2820
2870
2870
3280

1960
1960
2265
2265
2265

2515
2515
2515
2515
2515

13.9
13.9
16.4
16.4
18.7

696
696
883
883
826

407
407
545
545
540

1986
2236
3176
4016
5997

1692
1951
2851
3358
5625

CONTURA G2 

(opt. granite U-profile)

Upper part, X=1000, Y=1200, Z=600
Upper part, X=1000, Y=1600, Z=600
Upper part, X=1000, Y=2100, Z=600

2870
2845
3330

2265
2110
2110

2515
2515
2515

16.4
15.1
17.7

852
958
987

555
727
761

3092
3602
6084

2795
3372
5857

CONTURA G2 

(opt. granite U-profile)

Base, X=1000, Y=1200, Z=600
Base, X=1000, Y=1600, Z=600
Base, X=1000, Y=2100, Z=600

1805
2210
2515

1755
1755
1755

765
765
740

2.5
3.0
3.3

164
171
209

60
76
64

284
304
394

180
209
249

Pal box with accessories

1475

1070

1095

1.8

69

206

Controller

1100

1100

1100

1.4

90

227

Note:

All values specified above are maximum values. Any values not entered were not available at the time of printing. For 

technical reasons 

the

dimensions of the shipping crates (L, W and H) may deviate from the values specified here 

by as much as +50 mm

.

Summary of Contents for CONTURA G2

Page 1: ...Technical Services Installation Instructions CONTURA G2 Measuring Center ...

Page 2: ...ermissible air humidity 40 60 without condensation Sound level at installation site 90dB Max altitude above sea level for 100 125 V 3000 m for 220 240 V 2000 m Temperature Conditions for Operation of CMM The measuring uncertainty specified for your CMM can be guaranteed if 1 you use a measurement correction program and enter the workpiece and CMM temperatures as well as the temperature expansion c...

Page 3: ...erted by the CMM see the pertaining information in this bro chure 0 1 1 10 100 1000 0 10 20 30 40 50 X axis horizontal Y axis horizontal Z axis vertical Permissible foundation acceleration for Contura G2 w trigger probe and standard damping Acceleration values above the curve require an additional foundation Acceleration mm ss Frequency Hz ...

Page 4: ...0 10 20 30 40 50 X axis horizontal Y axis horizontal Z axis vertical Permissible foundation acceleration for Contura G2 measuring with standard damping Acceleration values above the curve require an additional foundation Acceleration mm ss Frequency Hz ...

Page 5: ... per taining national regulations also must be observed 2 3 Electric Power Supply Parameters To ensure trouble free data transmission between the CMM and the data station the line power supply must comply with the following specifications Line power supply Line voltage 1 N PE 100 110 115 120 125 230 240 VAC Max voltage fluctuations 10 Frequency 50 or 60 Hz Max frequency fluctuations 3 5 Max power ...

Page 6: ...e not obstructed by the base of the CMM 2 9 Transport Vehicles Please provide a forklift with sufficient carrying capacity and fork length to unload the packaged units and trans port them to the storage or installation site The loading capacities required of these vehicles are specified on the following pages All transport lanes and approaches should be kept open during the entire installation and...

Page 7: ...407 407 545 545 540 1986 2236 3176 4016 5997 1692 1951 2851 3358 5625 CONTURA G2 opt granite U profile Upper part X 1000 Y 1200 Z 600 Upper part X 1000 Y 1600 Z 600 Upper part X 1000 Y 2100 Z 600 2870 2845 3330 2265 2110 2110 2515 2515 2515 16 4 15 1 17 7 852 958 987 555 727 761 3092 3602 6084 2795 3372 5857 CONTURA G2 opt granite U profile Base X 1000 Y 1200 Z 600 Base X 1000 Y 1600 Z 600 Base X ...

Page 8: ... connection g CMM control connecting cable t Transport height without base X Y Z measuring range Length a Length b Length c Length t 700 700 600 1430 1525 920 1603 700 1000 600 1430 1830 1000 1200 600 1743 2030 1225 1000 1600 600 1743 2430 1000 2100 600 1743 2945 1756 4 CMM Installation Dimensions and Arrangement of Steel Plates 4 1 Installation Dimensions for CONTURA G2 all dimensions in mm scale...

Page 9: ... A to be installed by the customer e Telephone connection ISDN f CMM compressed air supply connection g CMM control connecting cable t Transport height without base CONTURA G2 with optional U profile X Y Z Measuring range Length a Length b Length c Length t 1000 1200 600 1743 2030 1200 1806 1000 1600 600 1743 2430 1000 2100 600 1743 2945 1959 Height of secured CMM assembly without pallet and witho...

Page 10: ...ding Admittance only for authorized personnel and may be entered only by persons who have received proper training operators min 1000 min 1000 min 1000 min 600 min 200 min 700 min 100 min 100 min 200 Floor marking or chain black and yellow d Controller c c The back door of the controller cabinet must be located at least 200mm away from the wall Moreover the air exchange with the surround ing envir...

Page 11: ...700 Y 1000 43 1 N cm2 CONTURA G2 X 1000 Y 1200 65 4 N cm2 X 1000 Y 1600 81 5 N cm2 X 1000 Y 2100 132 1 N cm2 CONTURA G2 w granite U table option X 1000 Y 1200 70 6 N cm2 X 1000 Y 1600 88 2 N cm2 X 1000 Y 2100 143 1 N cm2 Support A Cast iron plate Ø 150 included in delivery 4 4 Arrangement of the Steel Plates for Contura G2 The relative unevenness of the bearing surfaces must not exceed max 25mm 1 ...

Page 12: ... 4 700 237 700 127 89 1041 1 1 2 3 4 2 3 4 1321 1189 M12 16x Y X 256 456 656 856 Y grid position Bridge running surface Bridge drive X grid position 5 Dimensions of Granite Table and Fastening Hole Matrix 5 1 Measuring Table for Measuring Range of X 700 Y 700 Meas range ...

Page 13: ...8 744 8 944 8 1144 8 100 152 58 4 700 237 1000 127 89 1346 1 1 2 3 4 2 3 4 5 6 1626 1189 M12 24x Y X 256 456 656 856 Y grid position Bridge running surface Bridge drive X grid position 5 2 Measuring Table for Measuring Range of X 700 Y 1000 Meas range ...

Page 14: ...8 1144 8 1344 8 100 152 58 4 1000 237 1200 127 89 1545 1 1 2 3 4 5 6 2 3 4 5 6 7 1825 1492 M12 42x Y X 256 456 656 1056 856 1256 Y grid position Bridge running surface Bridge drive X grid position 5 3 Measuring Table for Measuring Range of X 1000 Y 1200 Meas range ...

Page 15: ...1344 8 1544 8 1744 8 100 152 58 4 1000 237 1600 127 89 1945 1 1 2 3 4 5 6 2 3 4 5 6 7 8 9 2225 1492 M12 54x Y X 256 456 656 1056 856 1256 Y grid position Bridge running surface Bridge drive X grid position 5 4 Measuring Table for Measuring Range of X 1000 Y 1600 Meas range ...

Page 16: ...4 8 1744 8 1944 8 2144 8 100 152 58 4 1000 237 2100 127 89 2464 1 1 2 3 4 5 6 2 3 4 5 6 7 8 9 2743 1494 M12 66x Y X 256 456 656 1056 856 1256 10 11 Y grid position Bridge running surface Bridge drive X grid position 5 5 Measuring Table for Measuring Range of X 1000 Y 2100 Meas range ...

Page 17: ... 1344 8 125 152 58 4 1000 237 1200 127 89 1545 1 1 2 3 4 5 6 2 3 4 5 6 7 1825 1492 M12 42x Y X 256 456 656 1056 856 1256 Y grid position Bridge running surface Bridge drive X grid position 5 6 Measuring Table with U Profile for Measuring Range of X 1000 Y 1200 Meas range ...

Page 18: ...544 8 1744 8 125 152 58 4 1000 237 1600 127 89 1945 1 1 2 3 4 5 6 2 3 4 5 6 7 8 9 2225 1492 M12 54x Y X 256 456 656 1056 856 1256 Y grid position Bridge running surface Bridge drive X grid position 5 7 Measuring Table with U Profile for Measuring Range of X 1000 Y 1600 Meas range ...

Page 19: ... 8 1944 8 2144 8 125 152 58 4 1000 237 2100 127 89 2464 1 1 2 3 4 5 6 2 3 4 5 6 7 8 9 2743 1494 M12 66x Y X 256 456 656 1056 856 1256 10 11 Y grid position Bridge running surface Bridge drive X grid position 5 8 Measuring Table with U Profile for Measuring Range of X 1000 Y 2100 Meas range ...

Page 20: ...Page 21 CONTURA G2 X 700 Y 1000 CONTURA G2 X 700 Y 700 CONTURA G2 X 1000 Y 1200 CONTURA G2 X 1000 Y 1600 CONTURA G2 X 1000 Y 2100 6 Cutout Patterns for Customer Floor Plan 6 1 Cutout Patterns for Contura G2 Scale 1 30 Controller ...