START-UP
Axiolab 5
Mounting standard components
ZEISS
05/2019
430037-7444-001
59
3.1.7
Mounting the Pol rotary stage
3.1.7.1
Removing the Pol rotary stage
•
Loosen the screw cap (Fig. 3-10/
6
) from the
spring housing (about three rotations).
•
Press the Pol rotary stage (Fig. 3-10/
4
) to the
front against the spring-loaded pin (Fig. 3-10/
7
),
lift it off the stage carrier (Fig. 3-10/
5
) from the
back and remove it upwards.
•
Re-tighten the screw cap (Fig. 3-10/
6
).
3.1.7.2
Attaching the Pol rotary stage
•
Where necessary, loosen the screw cap
(Fig. 3-10/
6
) of the spring housing with approx.
three rotations.
•
Place the Pol rotary stage with the groove of the
dovetail (stage bottom) on the spring-loaded
pin (Fig. 3-10/
7
).
•
Attach the rotary stage with the clamp screw
(Fig. 3-10/
8
) pointing to the front right.
•
Press the Pol rotary stage to the front against
the spring-loaded pin and lower it towards the
back into the stage carrier (Fig. 3-10/
5
), then
release it.
•
Re-tighten the screw cap (Fig. 3-10/
6
).
NOTE
The rotary stage must be mounted so that the vernier scale is on the left side and the clamp is
on the right side.
3.1.7.3
Dismantling the detachable specimen guide and mounting stage clips
•
Loosen the two clamp screws (Fig. 3-10/
1
) on the Pol specimen guide with an Allen wrench (AF 2).
Remove the Pol specimen guide (Fig. 3-10/
2
) by lifting it upwards.
•
Insert the stage clips (Fig. 3-10/
9
) into the holes provided.
Fig. 3-10
Replacing the snap-in Pol rotary
stage, detachable Pol specimen
guide and stage clips