PUTTING THE INSTRUMENT INTO OPERATION
Axio Vert.A1
Mounting Standard Components
Carl Zeiss
05/2012
431030-7044-001
57
3.1.11.5
Attaching the HAL 100 Illuminator
Before using the halogen lamp, be
sure to remove the replacement tool
for the halogen lamp from the
housing, as the tool might otherwise
get damaged by the heat generated.
(See Section 3.1.13.)
The halogen lamp can only be
operated with the external power
supply unit.
•
Loosen the clamping screw (Fig. 3-21/
2
) on the
reflected light port (Fig. 3-21/
1
) and remove the
cap.
•
Insert the illuminator housing (Fig. 3-21/
3
) with
the dovetail ring into the port (Fig. 3-21/
1
). Use
ball-headed screwdriver SW 3 and the locking
screw (Fig. 3-21/
2
) to fix the illuminator firmly
into position.
•
Insert the plug of the HAL 100 (Fig. 3-21/
5
) into
the
RL
socket at the back of the external power
supply unit (Fig. 3-21/
4
).
Fig. 3-21
Attaching the HAL 100 Illuminator