PUTTING THE INSTRUMENT INTO OPERATION
Axio Vert.A1
Mounting Optional Components
Carl Zeiss
05/2012
431030-7044-001
67
3.3.5
Changing the Beam-Splitting
Mirror in the Intermediate
Phototube
•
Put the existing slider (Fig. 3-33/
1
) of the
intermediate phototube on a desk.
•
Use the SW 2 wrench to loosen the screw,
remove the cover (Fig. 3-33/
2
) and the spring
3
).
•
Take out the existing beam-splitting mirror
(Fig. 3-33/
4
). Then insert the desired beam-
splitting mirror (with the coated surface
downward).
•
Put on the spring, so that the gib of the spring
engages into the existing groove and the bent
edge of the short sides around the mirror.
•
Put on the cover and fasten it with the screw. In
so doing, make sure that the opening sits on
the screw hole of the slider.
Fig. 3-33
Changing the Beam-Splitting
Mirror in the Intermediate
Phototube