CARE, MAINTENANCE, TROUBLESHOOTING AND SERVICE
ZEISS
Maintenance
Axio Observer
178
431004-7244-001
12/2016
6.2.3
Changing fuses on the
external power supply unit of
Axio Observer 7, 7 materials
Pull out the power plug before
changing fuses.
The Axio Observer 7, 7 materials is supplied with
voltage from the external power supply unit
VP232-2. The fuse compartment of the VP232-2
power supply unit is located at the rear of the unit
and contains two type T 4.0 A/H / 250 V fuses.
•
Pull out the power plug.
•
Pull out the fuse holder (Fig. 165/
2
) frontward.
•
Remove the fuses from the fuse holder and
insert new fuses.
•
Reinsert the fuse holder into the fuse
compartment (Fig. 165/
1
) until it clicks into
place.
•
Plug in the power plug.
Fig. 165
Changing fuses on the
external power supply unit for
Axio Observer 7, 7 materials