19
Changing the motor
AURA
®
handheld NIR - Rev. 2
EN_40_070_0030I | AURA
®
handheld NIR - Rev. 2 | Maintenance and Repair Manual
2
Remove the front panel and disconnect the cable
Note
NOTES ON DISASSEMBLY AND REASSEMBLY
It is much easier to remove and install the front panel if you also remove the
lamp holder (see “4 Replacing the halogen lamp” on page 14, step 1).
Take photos of the individual steps if necessary. This simplifies subsequent
work steps and avoids errors during reassembly.
Raise the front panel
1
enough to expose the cable underneath.
Then you can lay the front panel down slightly outside the housing.
Fig. 13
X800
X600
X704
X702*
1
2
3
Remove the front panel
Release the catch of the control cable connector in slot
X704
3
on the
“FBG PowerBoard” by pressing a narrow flat tip screwdriver into the small
opening on the side of the socket.
Proceed very carefully!
Disconnect the control cable from slot
X704
.
Release the catch of the motor cable connector in slot
X800
2
by pressing
on the connector latch.
Disconnect the control cable from slot
X800
.
Now remove the front panel completely.