14
Replacing the halogen lamp
AURA
®
handheld NIR - Rev. 2
Maintenance and Repair Manual | AURA
®
handheld NIR - Rev. 2 | EN_40_070_0030I
4 Replacing the halogen lamp
•
The display has been removed (see section “3 Changing the display”,
steps 1 to 6)
•
Put the following spare parts ready:
– Halogen lamp, order no. 000000-2345-437
•
Put the following tools and equipment ready:
– TX 8 and TX 10 screwdrivers
–
3 mm flat tip screwdriver
– Screw-holding tweezers
– Cotton gloves
1
Remove the lamp holder with halogen lamp
Unscrew the two Torx screws
2
from the base of the housing.
Use a TX 10 Torx screwdriver for this.
Carefully extract the lamp holder
1
upwards.
Disconnect the lamp cable
3
from the halogen lamp.
Fig. 8
2
3
1
Remove the lamp holder with halogen lamp
Requirements