8
Changing the display
AURA
®
handheld NIR - Rev. 2
Maintenance and Repair Manual | AURA
®
handheld NIR - Rev. 2 | EN_40_070_0030I
3 Changing the display
•
The control software of the device has been shut down and the device
switched off.
•
The ESD workstation and the ESD protective equipment have been made ready.
•
Put the following spare parts ready:
– Display, order no. 000000-2315-414
•
Put the following tools and equipment ready:
– Screwdriver TX 8
–
3 mm flat tip screwdriver
1
Remove battery
Procedure for device version 000000-2298-971:
Press in the two fasteners
1
on the battery compartment.
Carefully pull the cap
2
off the battery compartment.
Remove the battery
3
from the battery compartment using the grip strap.
Fig. 1
1
2
3
1
Remove battery (2298-971)
Close the battery compartment back up.