Zeiss Humphrey Systems Acuitus User Manual
PN 50648 Rev A June 2001
5
When you have made your selections, press the button under the
Page Down
icon to get to the
second setup screen where you can select printer and time and date options.
The second system setup screen (Model 5010 and 5015)
Note: Model 5000 does not have Print all PK data option.
Set Time
Press the Up Arrow and Down Arrow buttons until the correct time appears.
Next
will
take you am/pm, which you can set with the arrow buttons.
Set Date
Press the Up Arrow and Down Arrow buttons until you reach the correct date.
Time (Date)
Choose the time and date format you would like to use for your records: d = day,
m = month, y = year. You may also use a 24-hour clock, or the am/pm format.
Internal
The refractor’s internal printer can be turned
On
or
Off
with this option.
The default is
On
.
Paper
Type
Select
Standard
or
Label
(adhesive paper). The default is
Standard
.
Print Name/Date
Yes
tells the instrument to print a place for you to write the patient’s name on the
refraction printout. Both models print the date automatically.
No
omits name and date on
the printout. The default is
Yes
.
Print Multi-Ref
If you select
Yes
, the Acuitus will print the results of all of the multiple refractive
measurements it takes and print the average. If you select
No
, the
Acuitus
only prints the
average of its measurements.
Print Graphic
The graphic depicts the focal length of the eye and the position of the image relative to
the retina in a myopic, hyperopic, and/or astigmatic eye, depending on the patient’s
refraction. Select Yes or No.
When you have finished, you can press
Page Up
to go back to the first setup screen or
press
Return
. When you press
Return
, the last shown settings will become the new
defaults, and the Acuitus will take you to the start refraction screen.