The computer needs to be CE-marked and needs to satisfy the
requirements of DIN EN 60950 and/or DIN EN 60601-1.
Connecting the JMA-Optic system to a network/data pool can cause
unforeseen risks to the patient and third parties. If the database of the
software is installed in a network/data pool, the operator is
obliged to ascertain, analyze, evaluate and manage all of the associated
risks. In this context, the aspects of data protection, virus safety,
updates to the operating system and regular backups are of particular
importance. The risk assessments also have to include the subsequent
changes to the network/data pool, such as updates/upgrades to devices
and components that are connected with the network.
The connection of the PC system to the Internet must take place via
a professionally maintained IT network in conjunction with a hardware
firewall in order to minimize risks due to the Internet connection. Never
use the local computer with Internet access with administrator rights.
For the safe operation of the measuring system are the aspects of data
protection, virus security, updates of the operating system and regular
backups of the database on external data carriers of essential
importance and to be implemented by the user as measures of IT security.
Should problems with the hardware driver of the JMA-Optic system
occur then disconnect and restart the PC.
Installing a virus scanner on the PC can interfere with the proper
functioning of the application software. The manufacturer is not
reliable for damages caused by such programs.
Seite 30/69
Page 30/47
JMA-Optic
If the system is delivered without PC / laptop, please install the application software before
connecting the JMA-Optic to the computer. Please find information on the installation in the
user manual of the software.
6 Decommission the system
To take the system out of service, first close the software and then shut down
the PC. Move the face bow transport switch to position (0) and store face bow in transport
case. After 2 minutes at rest, the face bow automatically shuts off.
Finally, if necessary, remove the charger from the mains socket and store all components of
the JMA-Optic system in the carrying case.