background image

Zebra-Tech Ltd

  

 

D-OptoLogger Operation Manual

 

 

11

3.5 Calibration 

Under normal operating conditions, the D-Opto Logger should only require 
infrequent calibration. 
 
Calibration of the D-Opto Logger is a simple operation that can be carried out in 
the field. Due to the measurement principle of the D-Opto Logger, performing air 
calibrations are not advisable. The Zebra-Tech Do-Cal Kit greatly simplifies the 
field calibration of the D-Opto Logger. 
 
The offset and gain of the D-Opto Logger are calibrated individually. Normally 
only the offset will need to be calibrated as the gain is even more stable than the 
offset over time. The gain should only be calibrated 

AFTER

 the offset has been 

calibrated. 
 
When calibrating the D-Opto Logger, take care not to allow water to enter into the 
instrument. 
 

 Offset 

Calibration 

A reference solution of 0% dissolved oxygen is required for the offset 
calibration. This can be created by mixing a couple of teaspoons of 
sodium sulfite into 1 litre of distilled or fresh tap water.  
 
Procedure: 
1) Place the D-Opto Logger in the reference solution and allow 

equilibrating for at least 30 minutes.   

2) Power up the D-Opto Logger, and start D-OptoLog. Once the 

measurements have stabilised, press the “Calibrate” button in the main 
window. 

3)  Enter 0 (zero) in the offset calibration edit box and press the “Reset 

offset” button.  

4)  The D-Opto Logger performs a calibration routine followed by a self 

check; this can take up to 30 seconds.  Do not remove the D-Opto 
Logger from the reference solution during this period.  The operator will 
be notified when the process is complete.  

 

 Gain 

Calibration 

A solution of known dissolved oxygen content, other than 0%, should be 
used for the gain calibration. The most practical method is to create a fully 
aerated solution. This can be done by bubbling air through a container of 
distilled or fresh tap water, stirring frequently.  After around 30 minutes, 
the water can be assumed to be fully saturated.   
 
The actual dissolved oxygen saturation can be calculated by correcting for 
altitude, or preferably, the local barometric pressure if that is known.  The 
“Correction Calculator” can be used to calculate the corrected dissolved 
oxygen content of the fully aerated reference solution. As an example, if 

Summary of Contents for D-Opto

Page 1: ...D Opto Logger Dissolved Oxygen Logger Operation Manual...

Page 2: ...9 3 1 Installation 9 3 2 Sealing the D Opto Logger 9 3 3 Logging Endurance 9 3 4 Routine Maintenance 10 3 5 Calibration 11 Appendix 1 Data File Format 13 Appendix 2 Communication Cable Wiring Scheme 1...

Page 3: ...lved oxygen The basic principle involves subjecting a fluorescing compound ruthenium to a fixed wavelength of light When the light source is removed the compound emits a fluorescence the intensity of...

Page 4: ...on of the D OptoLog software During installation the user can select the working path and directory however it is recommended that the default is used c Program Files D OptoLog 2 2 Operation The D Opt...

Page 5: ...the computer s USB port via a USB to serial converter the port number will probably be either com port 4 or 5 The selected com port is maintained in an configuration file on the PC Once the correct p...

Page 6: ...the desired logging parameters The Plot button enables a graphical plot of historic D Opto Logger data to be displayed on the screen The Calibration button opens the D Opto Logger Calibration window F...

Page 7: ...nt of fully aerated water Figure 5 It will accept either altitude m or barometric pressure mBar By pressing the Restore defaults button the original factory calibration values are reinstated by the D...

Page 8: ...Zebra Tech Ltd D OptoLogger Operation Manual 8 Figure 5 Pressure Correction Calculator...

Page 9: ...ced prior underwater deployment 1 Using the O ring pick supplied remove the O rings from the inside of the end cap and the body of the D Opto Logger 2 Using a lint free tissue clean the O ring seats t...

Page 10: ...to allow a sufficient period of time for temperature equilibration to occur 2 Place the D Opto Logger in a solution of known dissolved oxygen content A solution of 0 dissolved oxygen saturation can be...

Page 11: ...1 Place the D Opto Logger in the reference solution and allow equilibrating for at least 30 minutes 2 Power up the D Opto Logger and start D OptoLog Once the measurements have stabilised press the Cal...

Page 12: ...minutes 2 Power up the D Opto Logger and start D OptoLog Once the measurements have stabilised press the Calibrate button in the main window to open the Calibration window 3 Enter the pressure correct...

Page 13: ...ptoLogger Operation Manual 13 Appendix 1 Data File Format Data Field Number Description 1 Year 2 Month 3 Day 4 Hour 5 Minute 6 Second 7 Battery voltage 8 Temperature Degrees C 9 Dissolved Oxygen 10 Di...

Page 14: ...Zebra Tech Ltd D OptoLogger Operation Manual 14 Appendix 2 Communication Cable Wiring Scheme...

Page 15: ...battery and communication socket there is an LED The LED flashes according to the current status of the D OptoLogger The interval between the flashing sequence may be up to 8 seconds 1 Flash Operation...

Page 16: ...e instrument has been carefully designed to prevent electrical interference impacting on the operation Under extreme conditions however noise may become a problem Track down the source of noise this c...

Page 17: ...ter Resolution Temperature 00 01 Deg C DO saturation 000 01 PPM 00 001 ppm Sensor Drift 1 per year provided measurement interval 10 minutes Power supply 1 x PP9 9 volt alkaline battery Duracell or Ene...

Page 18: ...1 983 97 343 1126 973 96 431 1413 963 95 519 1703 952 94 608 1995 942 93 698 2290 932 92 789 2587 922 91 880 2887 912 90 972 3190 902 89 1066 3496 899 88 1160 3804 882 87 1254 4115 871 86 1350 4430 86...

Page 19: ...Ltd D OptoLogger Operation Manual 19 Appendix 7 Useful Conversions Convert from To Calculation kPA mBar Multiply by 10 inHg mBar Multiply by 33 85 Feet Meters Multiply by 0 3048 Centigrade Fahrenheit...

Page 20: ...Zebra Tech Ltd D OptoLogger Operation Manual 20 Zebra Tech Ltd Contact Details PO Box 1668 Nelson New Zealand Tel 0064 03 5471590 Fax 0064 03 5471598 Email enquiry zebra tech co nz...

Reviews: