background image

 

How to select test times 

Testing times are dependent on factors such as part volume, test pressure, and leak rate. Other factors 
that will affect testing are things like compliance of the part under pressure or vacuum. Some amount of 
trial and error will be involved in finding acceptable values. However Zaxis has a few guidelines to follow in 
aid of your task. 

Fill 

– The time should be long enough such that the part reaches the test pressure and is stable. 

Settle

 – During the settle time the part is isolated from any incoming air. If the media of the part is 

susceptible to movement (compliance) the amount of pressure will drop slightly. The total settle time should 
be long enough to allow for the initial movement of the part to calm. 

Test

 – The test time is dependent on the needs of the product and the leak rate. Compare the 

decay of a good and bad part.  

 

2442 South 2570 West 

West Valley City, Utah 84117 

(801)264-1000 

www.zaxisinc.com 

 

 

 

P
R
E
S
S
U
R
E

 

T I M E

 

Fill

 

Test

 

Settle

Vent